Ovation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device delivered via catheter to treat descending abdominal aortic aneurysms (AAAs).
Publicado el 12 de noviembre de 2014 · Inicio del retiro: 20 de septiembre de 2014
Resumen
FDA registra el retiro Z-0125-2015 de dispositivos médicos de Clase I, publicado el 12 de noviembre de 2014 e iniciado el 20 de septiembre de 2014. Empresa responsable: Trivascular, Inc; Estado: Finalizado; Cantidad reportada: 0 Devices; Lote: The devices subject to this action are labeled for distribution outside of the United States; therefore, no UDI is on the device label or package. Model Number: TV-AB2980-D Lot Numbers: FS031814-11 FS052214-22 FS052814-36 FS040414-21 FS050614-56 FS040714-53 FS040514-02 FS042614-05 FS050614-53 FS032614-46 FS050214-27 FS050314-12 FS052214-48 FS052214-53 FS031814-44 FS031814-49 FS032714-04 FS032714-20 FS040414-30 FS040514-08 FS050114-01 FS050114-08 FS050514-18 FS050514-27 FS050614-54 FS050614-55 FS051914-32 FS051914-34 Expiry dates ranging "May-17" through "Jun-17". La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
TriVascular has identified select lots of Ovation Prime 29mm Aortic Body stent grafts that may be associated with rapid emptying of the fill polymer syringe when it is connected to the Autoinjector during the fill polymer injection step of the primary implant procedure.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Trivascular, Inc
- Distribuidor
- International Distribution Only -- Canada, Turkey, Austria, Ireland, Belgium, Italy, Switzerland, Spain, France, and Cyprus.
- Lote
- The devices subject to this action are labeled for distribution outside of the United States; therefore, no UDI is on the device label or package. Model Number: TV-AB2980-D Lot Numbers: FS031814-11 FS052214-22 FS052814-36 FS040414-21 FS050614-56 FS040714-53 FS040514-02 FS042614-05 FS050614-53 FS032614-46 FS050214-27 FS050314-12 FS052214-48 FS052214-53 FS031814-44 FS031814-49 FS032714-04 FS032714-20 FS040414-30 FS040514-08 FS050114-01 FS050114-08 FS050514-18 FS050514-27 FS050614-54 FS050614-55 FS051914-32 FS051914-34 Expiry dates ranging "May-17" through "Jun-17"
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDABaxter 2B8013 Empty INTRA VIA Container with PVC Ports (Sterile, nonpyrogenic fluid path) ,500 ml. Admixture12 de noviembre de 2014
- FDATerumo Custom Cardioplegia Delivery Set - Heat Exchanger Cardiopulmonary Bypass Item Numbers: 5852, 1601012 de noviembre de 2014
- FDABaxter 2B8011 Empty INTRA VIA Container with PVC Ports, Sterile fluid path, 150 ml, Admixture Products and12 de noviembre de 2014
- FDATerumo Custom Integrated Cardioplegia Delivery Set-LINE FOR 15501/16015/5852 Item Number: 16110 Product Usage:12 de noviembre de 2014
- FDAOvation Prime Abdominal Stent Graft System; Manufacturer: TriVascular, Inc. 3910 Brickway Blvd. Santa Rosa, CA2 de abril de 2014
- FDATriVascular brand Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European Model Number:21 de noviembre de 2012
Preguntas sobre este retiro
¿Ovation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device delivered via catheter to treat descending abdominal aortic aneurysms (AAAs). fue retirado del mercado?
Sí. FDA publicó el retiro Z-0125-2015 el 12 de noviembre de 2014. Peligro reportado: TriVascular has identified select lots of Ovation Prime 29mm Aortic Body stent grafts that may be associated with rapid emptying of the fill polymer syringe when it is connected to the Autoinjector during the fill polymer injection step of the primary implant procedure.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter