Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0156-2015

Terumo Custom Cardioplegia Delivery Set - Heat Exchanger Cardiopulmonary Bypass Item Numbers: 5852, 16010 Product Usage: This is a Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4:1 ratio. In includes a Conducer Heat Exchanger and a Bubble Trap module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity

Publicado el 12 de noviembre de 2014 · Inicio del retiro: 1 de marzo de 2011

Resumen

FDA registra el retiro Z-0156-2015 de dispositivos médicos de Clase II, publicado el 12 de noviembre de 2014 e iniciado el 1 de marzo de 2011. Empresa responsable: Terumo Cardiovascular Systems Corp; Estado: Finalizado; Cantidad reportada: 596 units; 876 units; Lote: Item Number: 58520-Lot Numbers: MF24, MM11, MN01. Manufacturing dates May 24, 20 I 0 through November I, 2010 Item Number: 1610-Lot Numbers: MM11, MN08, MNI5, MN29 MP07. Manufacturing dates October 11, 20 I 0 through December 7, 20 I 0. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Defect in the luer connector can compromise the seal on the female side of the connector and the compromised seal can introduce air in the table line

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Terumo Cardiovascular Systems Corp
Distribuidor
Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.
Lote
Item Number: 58520-Lot Numbers: MF24, MM11, MN01. Manufacturing dates May 24, 20 I 0 through November I, 2010 Item Number: 1610-Lot Numbers: MM11, MN08, MNI5, MN29 MP07. Manufacturing dates October 11, 20 I 0 through December 7, 20 I 0
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Terumo Custom Cardioplegia Delivery Set - Heat Exchanger Cardiopulmonary Bypass Item Numbers: 5852, 16010 Product Usage: This is a Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4:1 ratio. In includes a Conducer Heat Exchanger and a Bubble Trap module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity fue retirado del mercado?

Sí. FDA publicó el retiro Z-0156-2015 el 12 de noviembre de 2014. Peligro reportado: Defect in the luer connector can compromise the seal on the female side of the connector and the compromised seal can introduce air in the table line

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter