Terumo Custom Cardioplegia Delivery Set - Heat Exchanger Cardiopulmonary Bypass Item Numbers: 5852, 16010 Product Usage: This is a Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4:1 ratio. In includes a Conducer Heat Exchanger and a Bubble Trap module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity
Publicado el 12 de noviembre de 2014 · Inicio del retiro: 1 de marzo de 2011
Resumen
FDA registra el retiro Z-0156-2015 de dispositivos médicos de Clase II, publicado el 12 de noviembre de 2014 e iniciado el 1 de marzo de 2011. Empresa responsable: Terumo Cardiovascular Systems Corp; Estado: Finalizado; Cantidad reportada: 596 units; 876 units; Lote: Item Number: 58520-Lot Numbers: MF24, MM11, MN01. Manufacturing dates May 24, 20 I 0 through November I, 2010 Item Number: 1610-Lot Numbers: MM11, MN08, MNI5, MN29 MP07. Manufacturing dates October 11, 20 I 0 through December 7, 20 I 0. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Defect in the luer connector can compromise the seal on the female side of the connector and the compromised seal can introduce air in the table line
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Terumo Cardiovascular Systems Corp
- Distribuidor
- Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.
- Lote
- Item Number: 58520-Lot Numbers: MF24, MM11, MN01. Manufacturing dates May 24, 20 I 0 through November I, 2010 Item Number: 1610-Lot Numbers: MM11, MN08, MNI5, MN29 MP07. Manufacturing dates October 11, 20 I 0 through December 7, 20 I 0
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDATerumo Custom Cardioplegia Delivery Set- Catalog Number:165720 and Catalog Number: 6375 Product Usage:12 de noviembre de 2014
- FDATerumo Custom Integrated Cardioplegia Delivery Set-LINE FOR 15501/16015/5852 Item Number: 16110 Product Usage:12 de noviembre de 2014
- FDATerumo Custom Cardiovascular and Perfusion Procedure Kits Catalog Codes: Kits with a 5-digit code beginning12 de noviembre de 2014
- FDATerumo Cardioplegia Delivery Set- Item Number: 16130 Product Usage: The Cardioplegia Delivery Sets are12 de noviembre de 2014
- FDAOvation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime12 de noviembre de 2014
- FDABaxter 2B8013 Empty INTRA VIA Container with PVC Ports (Sterile, nonpyrogenic fluid path) ,500 ml. Admixture12 de noviembre de 2014
Preguntas sobre este retiro
¿Terumo Custom Cardioplegia Delivery Set - Heat Exchanger Cardiopulmonary Bypass Item Numbers: 5852, 16010 Product Usage: This is a Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4:1 ratio. In includes a Conducer Heat Exchanger and a Bubble Trap module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity fue retirado del mercado?
Sí. FDA publicó el retiro Z-0156-2015 el 12 de noviembre de 2014. Peligro reportado: Defect in the luer connector can compromise the seal on the female side of the connector and the compromised seal can introduce air in the table line
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter