FDAFDA · Alimentos y medicamentos·N.º oficial Z-0137-2015
Animas Vibe Insulin Infusion Pump and System. These products are indicated for continuous subcutaneous infusion of insulin for the treatment of insulin-requiring diabetes.
Publicado el 5 de noviembre de 2014 · Inicio del retiro: 25 de octubre de 2012
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Animas determined that keypad wear over time was contributing to an increased complaint rate associated with damaged keypads on the Vibe Insulin Infusion Pump and System. Animas has implemented an improvement to the keypad related to the material used to reduce susceptibility to damage and increase durability.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Animas Corporation
- Distribuidor
- Worldwide Distribution in the countries of France, Italy, Sweden, and UK..
- Lote
- Part Numbers 101200-57 100512-57 100510-02 100510-03 100510-04 100510-53 100510-56 100510-57 100510-63 100510-67 100511-02 100511-03 100511-04 100511-53 100511-56 100511-57 100511-63 100512-02 100512-02 100512-03 100512-04 100512-53 100512-56 100512-57 100512-63 100514-02 100514-03 100514-04 100514-53 100514-56 100514-57 100514-63 100515-02 100515-03 100515-04 100515-53 100515-56 100515-57 100515-63 101200-02 101002-03 101002-53 101002-56 101002-57 101002-63 101201-02 101201-03 101201-53 101210-56 101201-57 101201-63 101202-02 101202-03 101202-04 101202-53 101202-57 101202-63 101204-03 101204-04 101204-53 101204-57 101204-63 101205-03 101205-53 101205-56 101205-57 101205-63
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?