Merge Healthcare, Merge Hemo Programmable diagnostic computer. The system comprises the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC. The two units are connected via a serial interface. All vital parameters and evaluations are registered and calculated in the Patient Data Module. This data is then transmitted to the HeartSuite Hemodynamics Hemo Monitor PC via the serial interface. All data can be shown and monitored on the HeartSuite Hemodynamics Hemo Monitor PC. The Patient Data Module uses an internal battery charged from an external power input (RS 232/12V). The power supply, like the data transmission, is completely isolated from the visualization unit. The HeartSuite Hemodynamics Hemo Monitor PC is powered via the normal mains connection 230V/1 IOV.
Publicado el 5 de noviembre de 2014 · Inicio del retiro: 29 de septiembre de 2014
Resumen
FDA registra el retiro Z-0143-2015 de dispositivos médicos de Clase II, publicado el 5 de noviembre de 2014 e iniciado el 29 de septiembre de 2014. Empresa responsable: Merge Healthcare, Inc.; Estado: Finalizado; Cantidad reportada: 52; Lote: Merge Hemo 9.x with Nellcor SpO2. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
It has been reported that during use, the SpO2 value displayed on the Hemo Monitor may not update to reflect changes in the patient's oxygen value. It is also possible that if using the pulse rate of the SpO2 finger clip to calculate the patient's heart rate, it too may be subject to displaying a stale value.
Qué hacer
Voluntary: Firm initiated · E-Mail
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Merge Healthcare, Inc.
- Distribuidor
- Nationwide Distribution including the states of AZ, CA, FL, IL, IA, MA, MO, OH, SC, TN, and WI.
- Lote
- Merge Hemo 9.x with Nellcor SpO2
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAMerge PACS and Merge OrthoPACS, medical imaging diagnostic workstation2 de septiembre de 2026
- FDAMerge OrthoPACS, medical imaging diagnostic workstation12 de agosto de 2026
- FDAMerge PACS, medical imaging diagnostic workstation12 de agosto de 2026
- FDAVERICIS, Merge Cardio, Model/Catalog Number: Version 10.x, Software Version: 10.0, 10.1.1.1, 10.1.2, 10.2.0,26 de marzo de 2025
- FDADMI Duro-Med Industries Rollator Transport Chair; Wheel Size: 7 1/4"; Overall Width: 23 1/2"; Seat Height: 215 de noviembre de 2014
- FDAIMRIS Operating Room table (ORT200 and ORT300). Intended for use during diagnostic examinations or surgical5 de noviembre de 2014
Preguntas sobre este retiro
¿Merge Healthcare, Merge Hemo Programmable diagnostic computer. The system comprises the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC. The two units are connected via a serial interface. All vital parameters and evaluations are registered and calculated in the Patient Data Module. This data is then transmitted to the HeartSuite Hemodynamics Hemo Monitor PC via the serial interface. All data can be shown and monitored on the HeartSuite Hemodynamics Hemo Monitor PC. The Patient Data Module uses an internal battery charged from an external power input (RS 232/12V). The power supply, like the data transmission, is completely isolated from the visualization unit. The HeartSuite Hemodynamics Hemo Monitor PC is powered via the normal mains connection 230V/1 IOV. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0143-2015 el 5 de noviembre de 2014. Peligro reportado: It has been reported that during use, the SpO2 value displayed on the Hemo Monitor may not update to reflect changes in the patient's oxygen value. It is also possible that if using the pulse rate of the SpO2 finger clip to calculate the patient's heart rate, it too may be subject to displaying a stale value.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · E-Mail