FDAFDA · Alimentos y medicamentos·N.º oficial Z-0179-2015
Thyroglobulin (1D4) is a mouse anti-human monoclonal antibody produced as a tissue culture supernatant, and supplied in Tris buffered saline with carrier protein, containing 0.35% ProClin 950 as a preservative. Total volume = 7 mL. This reagent is for in vitro diagnostic use. Thyroglobulin (1D4) monoclonal antibody is intended to be used for the qualitative identification by light microscopy of thyroglobulin in formalin-fixed, paraffin-embedded tissue by immunohistochemical staining using an automated Bond system. The clinical interpretation of any staining or its absence should be complemented by morphological studies and proper controls and should be evaluated within the context of the patient's clinical history and other diagnostic tests by a qualified pathologist.
Publicado el 12 de noviembre de 2014 · Inicio del retiro: 18 de abril de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Bond Ready to Use Primary Antibody Thyroglobulin does not function as intended up to the expiry date on the product labelling. This was detected through an on-going stability program. There is a link between the age of the product and staining intensity.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Leica Microsystems, Inc.
- Distribuidor
- Worldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, MT, MY, NC, NH, NJ, NY, OH, ON, OR, PA, TN, TX, UT, VA, WA, WI and the District of Columbia, Internationally to Belgium, Bulgaria, France, Italy, Pakistan, Philippines, Russia, Saudi Arabia, Spain, Turkey, and the United Kingdom.
- Lote
- Product Code: PA0025; Lot Numbers and Expiration Dates: 24963, Expires 17 DEC 2014; 24228, Expires 15 OCT 2014; 23882, Expires 10 SEP 2014; 23366, Expires 23 JUL 2014
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States