Applied Medical's Inzii® 12/15mm Retrieval System. Applieds Inzii Tissue Retrieval Bag, Model Number CD004, is classified as a Class II device. It was cleared to market via 510(k) filing K060051. The retrieval bag is a disposable specimen retrieval bag intended for use as receptacle for the collection of tissue, organs, and calculi during laparoscopic procedures. Applieds CD004 has a shelf life of three years. It is provided sterile and packaged in a heat sealed Tyvek pouch.
Publicado el 20 de noviembre de 2013 · Inicio del retiro: 21 de octubre de 2013
Resumen
FDA registra el retiro Z-0183-2014 de dispositivos médicos de Clase II, publicado el 20 de noviembre de 2013 e iniciado el 21 de octubre de 2013. Empresa responsable: Applied Medical Resources Corp; Estado: Finalizado; Cantidad reportada: 29,506 units; Lote: Model Number CD004. Lot Nubmers: 1161268, 1161269, 1164934, 1168358, 1168361, 1169317, 1169318, 1170694, 1170700, 1170701, 1171214, 1172482, 1172664, 1174287, 1174793, 1174857, 1175123, 1175280, 1175472, 1176555, 1177023, 1177871, 1179069, 1179392, 1179460, 1179775, 1179890, 1180291, 1180411, 1180576, 1181647, 1181648, 1181649, 1181939, 1182760, 1182934, 1183225, 1183823, 1183957, 1184273, 1184728, 1184915, 1184916, 1187779, 1188817, 1189117, 1189957, 1190294, 1190358, 1191208, 1192057, 1192969, 1193196, 1193661, 1194010, 1194666, 1194807, 1195518, 1195727, 1195997, 1197182, 1197404, 1197832, 1198055, 1198968, 1199899, 1200658, 1200828, 1200829, 1200830, 1201739, 1201740, 1201741, 1203154, 1204054. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Applied Medical is conducting a voluntary recall of the Inzii® 12/15mm retrieval system. During shipment, the retrieval system packaging has the potential to become punctured with small holes, which could compromise the sterile barrier. The likelihood of this situation to occur and result in patient harm is highly unlikely; however, out of an abundance of caution for patient safety and a commitmen
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Applied Medical Resources Corp
- Distribuidor
- Worldwide Distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Lebanon, Malaysia, Mexico, New Zealand, Netherlands, Norway, Peru, Poland, Portugal, Saudi Arabia, Singapore, Spain, South Africa, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
- Lote
- Model Number CD004. Lot Nubmers: 1161268, 1161269, 1164934, 1168358, 1168361, 1169317, 1169318, 1170694, 1170700, 1170701, 1171214, 1172482, 1172664, 1174287, 1174793, 1174857, 1175123, 1175280, 1175472, 1176555, 1177023, 1177871, 1179069, 1179392, 1179460, 1179775, 1179890, 1180291, 1180411, 1180576, 1181647, 1181648, 1181649, 1181939, 1182760, 1182934, 1183225, 1183823, 1183957, 1184273, 1184728, 1184915, 1184916, 1187779, 1188817, 1189117, 1189957, 1190294, 1190358, 1191208, 1192057, 1192969, 1193196, 1193661, 1194010, 1194666, 1194807, 1195518, 1195727, 1195997, 1197182, 1197404, 1197832, 1198055, 1198968, 1199899, 1200658, 1200828, 1200829, 1200830, 1201739, 1201740, 1201741, 1203154, 1204054.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAEpix Universal Clip Applier, REF: CA500, STERILEEO3 de abril de 2024
- FDAEpix Universal Clip Applier, REF: CA500, STERILEEO included within the following kit Reference Numbers:3 de abril de 2024
- FDAClip Applier Kit, REF: K2887, UDI: (01)30607915139132(17)220822(30)1(10)3 de junio de 2020
- FDADR. THAMES - LAP CHOLE, REF: K0880, UDI: (01)306079151396203 de junio de 2020
- FDAProstiva RF Therapy Hand Piece Kit, REF 8929, Rx only, sterilized using Ethylene Oxide. The Prostiva RF Model20 de noviembre de 2013
- FDA2008K@ Home Blood lines Combiset True Flow Series for Hemodialysis use Catalog Number: 03-2962-3. Intended for20 de noviembre de 2013
Preguntas sobre este retiro
¿Applied Medical's Inzii® 12/15mm Retrieval System. Applieds Inzii Tissue Retrieval Bag, Model Number CD004, is classified as a Class II device. It was cleared to market via 510(k) filing K060051. The retrieval bag is a disposable specimen retrieval bag intended for use as receptacle for the collection of tissue, organs, and calculi during laparoscopic procedures. Applieds CD004 has a shelf life of three years. It is provided sterile and packaged in a heat sealed Tyvek pouch. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0183-2014 el 20 de noviembre de 2013. Peligro reportado: Applied Medical is conducting a voluntary recall of the Inzii® 12/15mm retrieval system. During shipment, the retrieval system packaging has the potential to become punctured with small holes, which could compromise the sterile barrier. The likelihood of this situation to occur and result in patient harm is highly unlikely; however, out of an abundance of caution for patient safety and a commitmen
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter