The Plum A+ Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+ Infusion System is designed to deliver parenteral, enteral, or epidural infusions over a broad range of infusion rates from multiple fluid container types. The Plum A+ Infusion System includes an infuser, disposable IV sets (hereafter called a set), optional accessories, and this operator's manual. The Plum A+ Infusion System with Hospira MedNet Software is indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.
Publicado el 20 de noviembre de 2013 · Inicio del retiro: 1 de febrero de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Plum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the administration set and indicates the presence of a full or partial distal occlusion. A broken distal pin can only be detected via a visual inspection of the distal pressure pin. A broken
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Hospira Inc.
- Distribuidor
- Worldwide Distribution - USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and U.S. Virgin Islands; and countries of: Costa Rica and Dominican Republic.
- Lote
- 1) List Number 11973; Serial Numbers: 12570148 , 12570150 , 12570154 , 12570163 , 12570170 , 12570177 , 12570204 , 12570211 , 12570212 , 12570255 , 12570270 , 12570319 , 12570359 , 12570385 , 12570413 , 12570425 , 12570455 , 12570487 , 12570662 , 12570695 , 12570854 , 12570915 , 12571080 , 12571243 , 12571437 , 12571631 , 12571692 , 12571776 , 12571826 , 12571835 , 12571837 , 12571842 , 12571877 , 12571880 , 12571886 , 12572317 , 12572369 , 12572586 , 12572909 , 12573547 , 12573567 , 12573837 , 12573990 , 12574424 , 12574501 , 12574524 , 12574556 , 12574599 , 12574619 , 12574731 , 12574749 , 12574853 , 12574858 , 12574860 , 12574865 , 12574961 , 12575626 , 12575698 , 12575886 , 12576125 , 12576443 , 12576485 , 12576518 , 12576593 , 12576609 , 12576710 , 12576722 , 12577271 , 12577685 , 12578856 , 12579183
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States