IO Fix IntraOsseous Fixation System; Extremity Medical. www.extremitymedical.com; Customer Service: 888.499.0079. The Extremity Medical Screw and Washer System is intended for reduction and internal fixation of arthrodeses, osteotomies, intra- and extrarticular fractures and nonunions of the small bones and joints of the foot, ankle, hand and wrist.
Publicado el 28 de noviembre de 2012 · Inicio del retiro: 13 de abril de 2011
Resumen
FDA registra el retiro Z-0344-2013 de dispositivos médicos de Clase II, publicado el 28 de noviembre de 2012 e iniciado el 13 de abril de 2011. Empresa responsable: Extremity Medical LLC; Estado: Finalizado; Cantidad reportada: 120 Units; Lote: Product coding is listed as Catalog Number: Lot Number and are as follows: 118-30014:107509; 118-30016:107510; 118-30018:107511; 118-30020:107512; 118-30020:107513; 118-30022:107514; 118-30024:107515; 118-30026:107516; 118-40020:107457; 118-40025:107458; 118-40030;107529; 118-40030;107459; 118-40035:107460; 118-40040:107461; 118-40045:107462; 118-40050:107463; 118-40320:107613; 118-40325:107614; 118-40330:107615; 118-40330:107616; 118-40335:107617; 118-40340:107618; 118-40345:107619; 118-40350:107620. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Product Event Reports revealed that lag screw passes through washer (e.g. X-Post) intra-operatively.
Qué hacer
Voluntary: Firm initiated · E-Mail
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Extremity Medical LLC
- Distribuidor
- Worldwide Distribution-USA (nationwide) including the states FL, MA, NC, CA, MD, VA, IN, AL, NJ, AZ, OK, MT, NY, OH, IA, MI, TX, WI, GA, IL, TN and NE, and the country of Switzerland.
- Lote
- Product coding is listed as Catalog Number: Lot Number and are as follows: 118-30014:107509; 118-30016:107510; 118-30018:107511; 118-30020:107512; 118-30020:107513; 118-30022:107514; 118-30024:107515; 118-30026:107516; 118-40020:107457; 118-40025:107458; 118-40030;107529; 118-40030;107459; 118-40035:107460; 118-40040:107461; 118-40045:107462; 118-40050:107463; 118-40320:107613; 118-40325:107614; 118-40330:107615; 118-40330:107616; 118-40335:107617; 118-40340:107618; 118-40345:107619; 118-40350:107620.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAExtremity Medical - Disposable Orthopedic Surgical Instrument - 1.6 mm Guidewire - Reference Number 101-0000627 de marzo de 2013
- FDAExtremity Medical, IO FiX System Accessory. Catalog Number 101-00010. This screw measuring device is an6 de febrero de 2013
- FDAIO FiX X-Post Reamer, Extremity Medical, www.extremitymedical.com; Customer Service: 888.499.0079 The X-Post5 de diciembre de 2012
- FDASuperstat Modified Collagen Hemostatic Sponge, 50mm Disc, 3.0%, Part# 9200-30. For use when hemostasis is28 de noviembre de 2012
- FDAUniCel Dxl 800 Access Immunoassay System, Part Numbers: 973100, A71456, and A71457. The UniCel DxI Immunoassay28 de noviembre de 2012
- FDASuperstat Modified Collagen Hemostatic Sponge, 32mm Disc, 3.0%, Part# 9100-30. For use when hemostasis is28 de noviembre de 2012
Preguntas sobre este retiro
¿IO Fix IntraOsseous Fixation System; Extremity Medical. www.extremitymedical.com; Customer Service: 888.499.0079. The Extremity Medical Screw and Washer System is intended for reduction and internal fixation of arthrodeses, osteotomies, intra- and extrarticular fractures and nonunions of the small bones and joints of the foot, ankle, hand and wrist. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0344-2013 el 28 de noviembre de 2012. Peligro reportado: Product Event Reports revealed that lag screw passes through washer (e.g. X-Post) intra-operatively.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · E-Mail