FDAFDA · Alimentos y medicamentos·N.º oficial Z-0419-2020
CoolBlue Hyper-Hypothermia System- body hypothermia system Model: Innercool - Product Usage: The system is composed of a heater, a compressor, a circulating pump and blankets/pads.
Publicado el 27 de noviembre de 2019 · Inicio del retiro: 2 de julio de 2019
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Warnings statements have been added- exceeding 40 degree C for extended periods may cause tissue damage. Additionally, a caution was added and clarifications were made regarding the use of the Automatic modes
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Cincinnati Sub-Zero Products LLC, a Gentherm Company
- Distribuidor
- Worldwide distribution - US Nationwide and countries of Afghanistan, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Belgium, Bhutan, Brazil, Brunei, Canada Chile, China, Colombia, Costa Rica, Ecuador, Egypt, El Salvador, France, Germany, Guatemala, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Lebanon, Malaysia, Mexico, Morocco, Nepal, Netherlands, New Zealand, Oman, Pakistan, Palestine, Panama, Philippines, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, Venezuela, Vietnam.
- Lote
- Serial Numbers: 073-CB-00001 through 162-CB-00294
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States