Cardiac Function Analysis Software Product Data No. MPDCT0301EAD CSCF-003A The software is applicable to the following CT systems: Aquilion ONE TSX-301C X-ray CT system with a 320-row 0.5 mm detector is provided with a 7.5-MHU large-capacity X-ray tube unit, and permits scanning over large areas at the high-speed of 0.35 s per rotation. The system features are; 160-mm-wide area detector, station, gantry and patient couch operation controls, image reconstruction, high-speed volume scan, high-quality images, selectable image slice thickness, dual-monitor system, improved image diagnosis functions, high-speed volume data workflow, exposure reduction, ECG-gated scan and reconstruction, and SUREFluoro. The CT cardiac function analysis software is used to analyze cardiac function using data obtained by performing cardiac MPR processing for ECG-gated scan data according to the cardiac axis. Aquilion ONE is a dynamic volume CT system that supports whole-body scanning. This 320-detector row system generates 640 slices per rotation using the coneXact reconstruction algorithm. In addition, the high-speed rotation mechanism and the fast reconstruction unit of the system allow the rapid image acquisition to further improve throughput in CT examinations.
Publicado el 3 de diciembre de 2014 · Inicio del retiro: 5 de noviembre de 2014
Resumen
FDA registra el retiro Z-0471-2015 de dispositivos médicos de Clase II, publicado el 3 de diciembre de 2014 e iniciado el 5 de noviembre de 2014. Empresa responsable: Toshiba American Medical Systems Inc; Estado: Finalizado; Cantidad reportada: 1 unit within the U.S. only; Lote: Item CFA Software Serial Number: Description TSX-301C/2D 2BA1362093 Aquilion One Vision. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Potential problem with the cardiac function analysis software (CFA). Incorrect analysis results may be displayed in the Function Parameters for the Entire Heart displayed as analysis results of the CFA and in a Left-Ventricular Volume Curve generated based on some of those parameters.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Toshiba American Medical Systems Inc
- Distribuidor
- US Nationwide Distribution to CA, MA, MD
- Lote
- Item CFA Software Serial Number: Description TSX-301C/2D 2BA1362093 Aquilion One Vision
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAINFX-8000C Fluoroscopic X-Ray Systems14 de marzo de 2018
- FDAAPLIO 500 TUS-A500 diagnostic ultrasound system19 de julio de 2017
- FDAXARIO 200 TUS-X200 diagnostic ultrasound system19 de julio de 2017
- FDAToshiba Kalare DREX-KL80 X-ray generator22 de marzo de 2017
- FDANon-sterile drape3 de diciembre de 2014
- FDACardiac Function Analysis Software Product Data No. MPDCT0301EAD CSCF-003A The software is applicable to the3 de diciembre de 2014
Preguntas sobre este retiro
¿Cardiac Function Analysis Software Product Data No. MPDCT0301EAD CSCF-003A The software is applicable to the following CT systems: Aquilion ONE TSX-301C X-ray CT system with a 320-row 0.5 mm detector is provided with a 7.5-MHU large-capacity X-ray tube unit, and permits scanning over large areas at the high-speed of 0.35 s per rotation. The system features are; 160-mm-wide area detector, station, gantry and patient couch operation controls, image reconstruction, high-speed volume scan, high-quality images, selectable image slice thickness, dual-monitor system, improved image diagnosis functions, high-speed volume data workflow, exposure reduction, ECG-gated scan and reconstruction, and SUREFluoro. The CT cardiac function analysis software is used to analyze cardiac function using data obtained by performing cardiac MPR processing for ECG-gated scan data according to the cardiac axis. Aquilion ONE is a dynamic volume CT system that supports whole-body scanning. This 320-detector row system generates 640 slices per rotation using the coneXact reconstruction algorithm. In addition, the high-speed rotation mechanism and the fast reconstruction unit of the system allow the rapid image acquisition to further improve throughput in CT examinations. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0471-2015 el 3 de diciembre de 2014. Peligro reportado: Potential problem with the cardiac function analysis software (CFA). Incorrect analysis results may be displayed in the Function Parameters for the Entire Heart displayed as analysis results of the CFA and in a Left-Ventricular Volume Curve generated based on some of those parameters.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter