Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0482-2021

ViewRay MRIdian Linac System: Model No. 20000, CE, Rated Supply Voltage - 380/480 VAC, 3 - Frequency Range (Hertz) - 50/60 Hz, Rated Input in Amperes - 210 A, UDI. The MRIdian system and the MRIdian Linac system, with magnetic resonance imaging capabilities, is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.

Publicado el 2 de diciembre de 2020 · Inicio del retiro: 20 de octubre de 2020

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

There is a potential that components of the receive coil can reach elevated temperatures which has the potential to cause a burn injury to the patient or user.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Viewray, Inc.
Distribuidor
Worldwide Distribution: US (nationwide) to states of: CA, NY, FL, IL, MI, MO, OR, WI and OUS countries of: France, Germany, Italy, United Kingdom, Taiwan, Netherlands, Israel, and Denmark.
Lote
All Lots/Serial Numbers
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta