Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0527-2021

CME America T-Syringe Pump - T34L PCA - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).

Publicado el 9 de diciembre de 2020 · Inicio del retiro: 30 de julio de 2020

Resumen

FDA registra el retiro Z-0527-2021 de dispositivos médicos de Clase II, publicado el 9 de diciembre de 2020 e iniciado el 30 de julio de 2020. Empresa responsable: CME America, LLC; Estado: Finalizado; Cantidad reportada: Total pumps = 4,182 (all models); Lote: Model T34L PCA - Part Number(s) (UDI Codes) 100-106PSL (UDI Code: B101100106PSL0) Serial Number: 26486 26485 26484 S93716 100-106PSLM (retired prior to UDI). La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
CME America, LLC
Distribuidor
Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.
Lote
Model T34L PCA - Part Number(s) (UDI Codes) 100-106PSL (UDI Code: B101100106PSL0) Serial Number: 26486 26485 26484 S93716 100-106PSLM (retired prior to UDI)
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿CME America T-Syringe Pump - T34L PCA - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site). fue retirado del mercado?

Sí. FDA publicó el retiro Z-0527-2021 el 9 de diciembre de 2020. Peligro reportado: Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter