Abbott Healon EndoCoat Ophthalmic Viscosurgical Device 3% sodium hyaluronate 0.85 mL 30 mg/mL Sterile A Solution Sterile EO Packaging and Cannula Product of USA Abbott Healon Duet Dual Pack 0.55 ml OVD 0.85 ml Healon EndoCoat OVD Contents: -one sterile syringe of Healon Ophthalmic Viscosurgical Device (OVD) 0.55 ml (10 mg/ml Sodium Hyaluronate) -one sterile syringe of Healon EndoCoat Ophthalmic Viscosurgical Device (OVD) 0.85 ml (30 mg/ml Sodium Hyaluronate) Part Number: US: VT585U and 57502000 (packaged in Healon Duet 10290080) OUS: VT585 and 57550500 (packaged in Healon duet 10220010, 10220011 and 10220012)
Publicado el 17 de diciembre de 2014 · Inicio del retiro: 6 de noviembre de 2014
Resumen
FDA registra el retiro Z-0530-2015 de dispositivos médicos de Clase II, publicado el 17 de diciembre de 2014 e iniciado el 6 de noviembre de 2014. Empresa responsable: Abbott Medical Optics Inc (AMO); Estado: Finalizado; Cantidad reportada: 484,881 Nationwide (U.S. and Puerto Rico) and 483,984 Internationally; Lote: Part Number: US: VT585U and 57502000 (packaged in Healon Duet 10290080) OUS: VT585 57550500 (packaged in Healon Duet 10220010, 10220011, and 10220012) Healon EndoCoat Device identification: For Healon EndoCoat, each lot is assigned a unique lot number for traceability. The lot number is a 6-digit numerical code that is equal to the manufacturing order number assigned. The numbers 1-6 are sequentially increased representing the manufacturing order number. Example: Lot 024992. For Healon EndoCoat packaged in Healon Duet, each Healon Duet final product is identified witha unique batch number. The batch number consists of two alphabetic characters followed by five numeric characters. The first alphabetic character of each batch number designates the manufacturing site. The second alphabetic character of each batch number shall be sequential numbers starting with 30001. Example: Lot UP30729 (This lot number would represent the 729th unit manufactured by AMO in 2014). La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
AMO has received complaints where the finger grip dislodged while depressing the syringe plunger. The finger grip dislodgement could result in the surgeon's hand unintentionally moving the cannula further than intended, resulting in potential trauma to the eye.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Abbott Medical Optics Inc (AMO)
- Distribuidor
- Worldwide Distribution - USA (Nationwide and Puerto Rico) and internationally to the following countries; Italy, Austria, Saudi Arabia, Turkey, Hong Kong, Canada, Germany, Australia, Columbia, Venezuela, Netherlands, Portugal, New Zealand, Denmark, Argentina, France, Belgium, Costa Rica, Singapore, Ukraine, Guadeloupe, Kuwait, Chile, Peru, Mexico, Czech Republic, Serbia, Hungary, Slovenia, Guatemala, Finland, Honduras, Malaysia, United Kingdom, Spain, Switzerland, Israel, and Croatia.
- Lote
- Part Number: US: VT585U and 57502000 (packaged in Healon Duet 10290080) OUS: VT585 57550500 (packaged in Healon Duet 10220010, 10220011, and 10220012) Healon EndoCoat Device identification: For Healon EndoCoat, each lot is assigned a unique lot number for traceability. The lot number is a 6-digit numerical code that is equal to the manufacturing order number assigned. The numbers 1-6 are sequentially increased representing the manufacturing order number. Example: Lot 024992. For Healon EndoCoat packaged in Healon Duet, each Healon Duet final product is identified witha unique batch number. The batch number consists of two alphabetic characters followed by five numeric characters. The first alphabetic character of each batch number designates the manufacturing site. The second alphabetic character of each batch number shall be sequential numbers starting with 30001. Example: Lot UP30729 (This lot number would represent the 729th unit manufactured by AMO in 2014).
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDASoft Contact Lens17 de diciembre de 2014
- FDAThe Affinity Fusion Oxygenator with Integrated Arterial Filter and Carmeda Bioactive Surface is a single-use,17 de diciembre de 2014
- FDATrevo Pro 4 Stent Retriever, Model number 80021; U.S. Indications for use: The Trevo Retriever is intended to17 de diciembre de 2014
- FDAA uterine tenaculum is a hook-like instrument used to seize and hold the cervix or fundus sold under the17 de diciembre de 2014
- FDAAlcon Single-Use Curved Jewelers/Iris Bipolar Forceps REF 8065129101 Sterile RX only; Alcon Single-Use17 de diciembre de 2014
- FDAAMO Sensar Intraocular Lenses (IOLs), Model AR40e, Sensar Foldable IOL with OptiEdge Product Usage: These are20 de marzo de 2013
Preguntas sobre este retiro
¿Abbott Healon EndoCoat Ophthalmic Viscosurgical Device 3% sodium hyaluronate 0.85 mL 30 mg/mL Sterile A Solution Sterile EO Packaging and Cannula Product of USA Abbott Healon Duet Dual Pack 0.55 ml OVD 0.85 ml Healon EndoCoat OVD Contents: -one sterile syringe of Healon Ophthalmic Viscosurgical Device (OVD) 0.55 ml (10 mg/ml Sodium Hyaluronate) -one sterile syringe of Healon EndoCoat Ophthalmic Viscosurgical Device (OVD) 0.85 ml (30 mg/ml Sodium Hyaluronate) Part Number: US: VT585U and 57502000 (packaged in Healon Duet 10290080) OUS: VT585 and 57550500 (packaged in Healon duet 10220010, 10220011 and 10220012) fue retirado del mercado?
Sí. FDA publicó el retiro Z-0530-2015 el 17 de diciembre de 2014. Peligro reportado: AMO has received complaints where the finger grip dislodged while depressing the syringe plunger. The finger grip dislodgement could result in the surgeon's hand unintentionally moving the cannula further than intended, resulting in potential trauma to the eye.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter