Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0528-2015

Trevo Pro 4 Stent Retriever, Model number 80021; U.S. Indications for use: The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Publicado el 17 de diciembre de 2014 · Inicio del retiro: 13 de noviembre de 2013

Resumen

FDA registra el retiro Z-0528-2015 de dispositivos médicos de Clase II, publicado el 17 de diciembre de 2014 e iniciado el 13 de noviembre de 2013. Empresa responsable: Concentric Medical Inc; Estado: Finalizado; Cantidad reportada: 10 units; Lote: Lot number/Expiration date: Lot Number 36774: March 2015 Lot Number 36432: December 2014 Lot Number 36757: March 2015 Lot Number 36761: March 2015. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Product shipped to US customers had non U.S. Instruction for use with indication for Use that were not aligned with U.S. indications.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Concentric Medical Inc
Distribuidor
US Distribution to states of: CA, IN, PA, AL, and NJ.
Lote
Lot number/Expiration date: Lot Number 36774: March 2015 Lot Number 36432: December 2014 Lot Number 36757: March 2015 Lot Number 36761: March 2015
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Trevo Pro 4 Stent Retriever, Model number 80021; U.S. Indications for use: The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0528-2015 el 17 de diciembre de 2014. Peligro reportado: Product shipped to US customers had non U.S. Instruction for use with indication for Use that were not aligned with U.S. indications.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter