Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL
Publicado el 4 de diciembre de 2024 · Inicio del retiro: 9 de octubre de 2024
Resumen
FDA registra el retiro Z-0554-2025 de dispositivos médicos de Clase II, publicado el 4 de diciembre de 2024 e iniciado el 9 de octubre de 2024. Empresa responsable: Shimadzu Corp. Analytical & Measuring Instruments Div.; Estado: En curso; Cantidad reportada: 48; Lote: Model/UDI-DI/Serial Numbers: SIL-40C XR CL/04540217420582/L2C316051002, L2C316051003, L2C316051004, L2C316051005, L2C316051006, L2C316051007, L2C316151012, L2C316151013, L2C316151014, L2C316151015. SIL-40C X3 CL/04540217420575/L2C305850001, L2C305950002, L2C306151025, L2C306051001, L2C306051002, L2C306051003, L2C306051004, L2C306051005, L2C306051007, L2C306051006, L2C306051008, L2C306051009, L2C306051010, L2C306051011, L2C306051012, L2C306051013, L2C306051014, L2C306051015, L2C306151017, L2C306151018, L2C306151022, L2C306151023, L2C306151024, L2C306151026, L2C306151028, L2C306151027, L2C306151029, L2C306151030, L2C306151031, L2C306151032, L2C306151033, L2C306151034, L2C306151035, L2C306151036, L2C306151037, L2C306151038, L2C306151040, L2C306151039. Firmware versions below 1.16. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Shimadzu Corp. Analytical & Measuring Instruments Div.
- Distribuidor
- Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE.
- Lote
- Model/UDI-DI/Serial Numbers: SIL-40C XR CL/04540217420582/L2C316051002, L2C316051003, L2C316051004, L2C316051005, L2C316051006, L2C316051007, L2C316151012, L2C316151013, L2C316151014, L2C316151015. SIL-40C X3 CL/04540217420575/L2C305850001, L2C305950002, L2C306151025, L2C306051001, L2C306051002, L2C306051003, L2C306051004, L2C306051005, L2C306051007, L2C306051006, L2C306051008, L2C306051009, L2C306051010, L2C306051011, L2C306051012, L2C306051013, L2C306051014, L2C306051015, L2C306151017, L2C306151018, L2C306151022, L2C306151023, L2C306151024, L2C306151026, L2C306151028, L2C306151027, L2C306151029, L2C306151030, L2C306151031, L2C306151032, L2C306151033, L2C306151034, L2C306151035, L2C306151036, L2C306151037, L2C306151038, L2C306151040, L2C306151039. Firmware versions below 1.16.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Japan
Retiros relacionados
- FDAAutosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL4 de diciembre de 2024
- FDATempus LS-Manual Defibrillator Model 00-3020 The Tempus LS-Manual is a defibrillator that can deliver a shock4 de diciembre de 2024
- FDAMerlin@home with the MerlinOnDemand capability Transmitter REF EX1150 Merlin@home with the MerlinOnDemand"4 de diciembre de 2024
- FDAENDO KIT, Medline Kit SKU DYKE1577A, Component Number 57469, 10 per case; Medline Convenience Kits containing4 de diciembre de 2024
- FDAMerlin@home Transmitter REF EX1100 The Merlin@home" (M@h) transmitter is intended to transfer stored data from4 de diciembre de 2024
- FDABaxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number 5C4482; use in Peritoneal Dialysis4 de diciembre de 2024
Preguntas sobre este retiro
¿Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL fue retirado del mercado?
Sí. FDA publicó el retiro Z-0554-2025 el 4 de diciembre de 2024. Peligro reportado: There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter