Secure II Med/Surg Bed Secure II and the 3002 S3 are the beds typically found in the Med-Surg units of the hospitals. The bed is designed for a 15 year expected service life under normal use conditions, and with appropriate periodic maintenance as described in the maintenance manual for each device. There are minimal differences between the functionality of the two models, each of which function as a highly complex system of a large number of individual components.

Publicado el 8 de enero de 2014 · Inicio del retiro: 26 de noviembre de 2013

Resumen

FDA registra el retiro Z-0617-2014 de dispositivos médicos de Clase II, publicado el 8 de enero de 2014 e iniciado el 26 de noviembre de 2013. Empresa responsable: Stryker Medical Division of Stryker Corporation; Estado: Finalizado; Cantidad reportada: 108,769 Total; Lote: Models: 3002 EX, 3002 KCI, 3002 CUSTOM, 3002 ZOOM, 3002000000, and 3002 ZX-SB. And Serial Numbers: 060215711 , 060215712 , 060215713 , 060215714 , 060215715 , 060215716 , 060215717 , 060215718 , 060215719 , 060215720 , 060215721 , 060215722 , 060215723 , 060215724 , 060215725 , 060215726 , 060215727 , 060215728 , 060215729 , 060215730 , 060215731 , 060215732 , 060215733 , 060215734 , 060215735 , 060215736 , 060215737 , 060215738 , 060215739 , 060215740 , 060215741 , 060215742 , 060215743 , 060215744 , 060215761 , 060215762 , 060215763 , 060215764 , 060215765 , 060215766 , 060215767 , 060215768 , 060215769 , 060215770 , 060215771 , 060215772 , 060215773 , 060215774 , 060215775 , 060215776 , 060215777 , 060215778 , 060215779 , 060215780 , 060215781 , 060215782 , 060215783 , 060215784 , 060215785 , 060215786 , 050916199 , 060515187 , 080615291 , 080615292 , 080615293 , 080615294 , 080615295 , 080615296 , 080615297 , 080615298 , 080615299 , 080615300 , 080615301 , 080615302 , 080615303 , 080615304 , 080615305 , 080615306 , 080615307 , 080615308 , 080615309 , 080615310 , 080615311 , 080615312 , 050815989 , 050815990 , 050815991 , 050815992 , 050815993 , 050816164 , 050816165 , 050816166 , 050816167 , 050816168 , 050816169 , 050816170 , 050816171 , 050816172 , 050816173 , 050816174 , 050816175 , 050816176 , 050816177 , 050816178 , 050816179 , 050816180 , 050816181 , 050816182 , 050816183 , 050816184 , 050816185 , 050816186 , 050816187 , 050816188 , 050816189 , 050816190 , 050816191 , 050816192 , 050816193 , 050816194 , 050816195 , 050816196 , 050816197 , 050816198 , 050816199 , 050816200 , 050816201 , 050816202 , 050816203 , 050816204 , 050816205 , 050816206 , 050816207 , 050816208 , 050816443 , 050816444 , 050816445 , 050816446 , 050816447 , 050816448 , 050816449 , 050816450 , 050816451 , 050816452 , 050816453 , 050816454 , 050816455 , 050816456 , 050816457 , 050816458 , 050816460 , 050816461 , 050816462 , 050816463 , 050816464 , 050816465 , 050816466 , 050816467 , 051015830 , 051015831 , 051015832 , 051015833 , 051015834 , 051015835 , 051015836 , 051015837 , 051015838 , 051015839 , 051015840 , 051015841 , 051015842 , 051015843 , 051015844 , 051015845 , 051015846 , 051015847 , 051015848 , 051015849 , 051015850 , 051015851 , 051015852 , 051015853 , 051015854 , 051015855 , 051015856 , 051015857 , 060915123 , 060915124 , 060915125 , 060915126 , 070416726 , 070817045 , 080215003 , 080215004 , 080215007 , 080215011 , 080215012 , 080415722 , 080415731 , 080615855 , 090515943 , 090515945 , 090515992 , 090515993 , 090515994 , 090515995 , 090515997 , 090515999 , 090516000 , 090516001 , 090516002 , 090516004 , 090516005 , 090516008 , 090516010 , 090516014 , 070217581 , 070816127 , 070816130 , 070816132 , 070817054 , 070817055 , 070817056 , 070817058 , 070817060 , 070817063 , 070817065 , 070817066 , 070817067 , 070817068 , 070817069 , 070817073 , 070817074 , 070915523 , 070916268 , 070916282 , 070916306 , 071015109 , 071015127 , 071217136 , 071217137 , 080215220 , 080215222 , 080215227 , 080216899 , 060615918 , 050516142 , 050516143 , 050516144 , 050516145 , 050516146 , 050516147 , 050516148 , 050516149 , 050516150 , 050516151 , 050516152 , 050516153 , 050516154 , 050516155 , 050516156 , 050516157 , 050516158 , 050516159 , 050516160 , 050516161 , 050516162 , 050516163 , 050516164 , 050516165 , 050516166 , 050516167 , 050516168 , 050516169 , 050516170 , 050516171 , 050616204 , 050616210 , 05061621. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Stryker Medical has identified a potential issue associated with Secure II and S3 Med/Surg beds built between April 2nd, 2005 and July 26th, 2011. The affected beds have the potential to experience damage to the Foot End Cover and/or Power Coil Cable due to reduced clearance between the Foot End Cover and Foot End Lift Header.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Stryker Medical Division of Stryker Corporation
Distribuidor
Worldwide Distribution - USA (nationwide) and Internationally to Canada, Chile, China, Argentina, Brazil, Germany, Korea, Mexico, Singapore, UK, Europe, Middle East, Africa, and Latin America.
Lote
Models: 3002 EX, 3002 KCI, 3002 CUSTOM, 3002 ZOOM, 3002000000, and 3002 ZX-SB. And Serial Numbers: 060215711 , 060215712 , 060215713 , 060215714 , 060215715 , 060215716 , 060215717 , 060215718 , 060215719 , 060215720 , 060215721 , 060215722 , 060215723 , 060215724 , 060215725 , 060215726 , 060215727 , 060215728 , 060215729 , 060215730 , 060215731 , 060215732 , 060215733 , 060215734 , 060215735 , 060215736 , 060215737 , 060215738 , 060215739 , 060215740 , 060215741 , 060215742 , 060215743 , 060215744 , 060215761 , 060215762 , 060215763 , 060215764 , 060215765 , 060215766 , 060215767 , 060215768 , 060215769 , 060215770 , 060215771 , 060215772 , 060215773 , 060215774 , 060215775 , 060215776 , 060215777 , 060215778 , 060215779 , 060215780 , 060215781 , 060215782 , 060215783 , 060215784 , 060215785 , 060215786 , 050916199 , 060515187 , 080615291 , 080615292 , 080615293 , 080615294 , 080615295 , 080615296 , 080615297 , 080615298 , 080615299 , 080615300 , 080615301 , 080615302 , 080615303 , 080615304 , 080615305 , 080615306 , 080615307 , 080615308 , 080615309 , 080615310 , 080615311 , 080615312 , 050815989 , 050815990 , 050815991 , 050815992 , 050815993 , 050816164 , 050816165 , 050816166 , 050816167 , 050816168 , 050816169 , 050816170 , 050816171 , 050816172 , 050816173 , 050816174 , 050816175 , 050816176 , 050816177 , 050816178 , 050816179 , 050816180 , 050816181 , 050816182 , 050816183 , 050816184 , 050816185 , 050816186 , 050816187 , 050816188 , 050816189 , 050816190 , 050816191 , 050816192 , 050816193 , 050816194 , 050816195 , 050816196 , 050816197 , 050816198 , 050816199 , 050816200 , 050816201 , 050816202 , 050816203 , 050816204 , 050816205 , 050816206 , 050816207 , 050816208 , 050816443 , 050816444 , 050816445 , 050816446 , 050816447 , 050816448 , 050816449 , 050816450 , 050816451 , 050816452 , 050816453 , 050816454 , 050816455 , 050816456 , 050816457 , 050816458 , 050816460 , 050816461 , 050816462 , 050816463 , 050816464 , 050816465 , 050816466 , 050816467 , 051015830 , 051015831 , 051015832 , 051015833 , 051015834 , 051015835 , 051015836 , 051015837 , 051015838 , 051015839 , 051015840 , 051015841 , 051015842 , 051015843 , 051015844 , 051015845 , 051015846 , 051015847 , 051015848 , 051015849 , 051015850 , 051015851 , 051015852 , 051015853 , 051015854 , 051015855 , 051015856 , 051015857 , 060915123 , 060915124 , 060915125 , 060915126 , 070416726 , 070817045 , 080215003 , 080215004 , 080215007 , 080215011 , 080215012 , 080415722 , 080415731 , 080615855 , 090515943 , 090515945 , 090515992 , 090515993 , 090515994 , 090515995 , 090515997 , 090515999 , 090516000 , 090516001 , 090516002 , 090516004 , 090516005 , 090516008 , 090516010 , 090516014 , 070217581 , 070816127 , 070816130 , 070816132 , 070817054 , 070817055 , 070817056 , 070817058 , 070817060 , 070817063 , 070817065 , 070817066 , 070817067 , 070817068 , 070817069 , 070817073 , 070817074 , 070915523 , 070916268 , 070916282 , 070916306 , 071015109 , 071015127 , 071217136 , 071217137 , 080215220 , 080215222 , 080215227 , 080216899 , 060615918 , 050516142 , 050516143 , 050516144 , 050516145 , 050516146 , 050516147 , 050516148 , 050516149 , 050516150 , 050516151 , 050516152 , 050516153 , 050516154 , 050516155 , 050516156 , 050516157 , 050516158 , 050516159 , 050516160 , 050516161 , 050516162 , 050516163 , 050516164 , 050516165 , 050516166 , 050516167 , 050516168 , 050516169 , 050516170 , 050516171 , 050616204 , 050616210 , 05061621
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Secure II Med/Surg Bed Secure II and the 3002 S3 are the beds typically found in the Med-Surg units of the hospitals. The bed is designed for a 15 year expected service life under normal use conditions, and with appropriate periodic maintenance as described in the maintenance manual for each device. There are minimal differences between the functionality of the two models, each of which function as a highly complex system of a large number of individual components. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0617-2014 el 8 de enero de 2014. Peligro reportado: Stryker Medical has identified a potential issue associated with Secure II and S3 Med/Surg beds built between April 2nd, 2005 and July 26th, 2011. The affected beds have the potential to experience damage to the Foot End Cover and/or Power Coil Cable due to reduced clearance between the Foot End Cover and Foot End Lift Header.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter