Greatbatch Stiffer Coaxial Micro-Introducer. Model #/ Name: 10904-001 Stiffer Coaxial Micro-Introducer 4FR; 10904-002 Stiffer Coaxial Micro-Introducer 5FR. The radiopaque Stiffer Coaxial Micro-Introducer (Figure 1) consists of a dilator and sheath, available in 4FR and 5FR. The inner dilator includes a stainless steel stiffening hypotube. The Coaxial Micro-Introducer is indicated for percutaneous introduction of up to a 0.038 in. guidewire or catheter into the vascular system through an initial puncture of a 21G introducer. The device is sold to customers in a bulk non-sterile configuration.
Publicado el 30 de noviembre de 2016 · Inicio del retiro: 8 de septiembre de 2016
Resumen
FDA registra el retiro Z-0623-2017 de dispositivos médicos de Clase II, publicado el 30 de noviembre de 2016 e iniciado el 8 de septiembre de 2016. Empresa responsable: Greatbatch Medical; Estado: Finalizado; Cantidad reportada: 23,075; Lote: GBP3176925, GBP3176927, VN3470927, VN3470929, VN3486000, VN3485998, VN3495047, GBP3421664, VN3433259, VN3451654, VN3487654, VN3491069, VN3470925, VN3451646. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Greatbatch Medical has identified an issue with the Stiffer Coaxial Micro-Introducer 4FR and 5FR after a field complaint reported that the rotating luer detached from the dilator hub during use.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Greatbatch Medical
- Distribuidor
- US: NY, TX.
- Lote
- GBP3176925, GBP3176927, VN3470927, VN3470929, VN3486000, VN3485998, VN3495047, GBP3421664, VN3433259, VN3451654, VN3487654, VN3491069, VN3470925, VN3451646.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDABiosense Webster MobiCath Bi-Directional Guiding Sheath, Model D140010 (small curve) and D140011 (large9 de agosto de 2017
- FDAViaPeel PTFE Peelable Introducer 5FR, Part Number 10890-0069 de agosto de 2017
- FDALinear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159) and EZ17 de mayo de 2017
- FDAPhilips Healthcare System X-ray Tomography Computed :30 de noviembre de 2016
- FDA3.0 mm Guide Rod/950 mm with Smooth Tip, sterile30 de noviembre de 2016
- FDATorque Limiting Screwdriver: Greatbatch Medical, Torque Limiting Driver, Non Sterile, Rx Only. PRECIMED, Non8 de junio de 2016
Preguntas sobre este retiro
¿Greatbatch Stiffer Coaxial Micro-Introducer. Model #/ Name: 10904-001 Stiffer Coaxial Micro-Introducer 4FR; 10904-002 Stiffer Coaxial Micro-Introducer 5FR. The radiopaque Stiffer Coaxial Micro-Introducer (Figure 1) consists of a dilator and sheath, available in 4FR and 5FR. The inner dilator includes a stainless steel stiffening hypotube. The Coaxial Micro-Introducer is indicated for percutaneous introduction of up to a 0.038 in. guidewire or catheter into the vascular system through an initial puncture of a 21G introducer. The device is sold to customers in a bulk non-sterile configuration. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0623-2017 el 30 de noviembre de 2016. Peligro reportado: Greatbatch Medical has identified an issue with the Stiffer Coaxial Micro-Introducer 4FR and 5FR after a field complaint reported that the rotating luer detached from the dilator hub during use.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter