Medical Surgical Bed, Model 3005S3EX Medical Surgical Bed The S3" MedSurg Bed, Model 3005 is intended to support and transport patients within the Med/Surg and Critical Care hospital environments. The bed is typically used in pre-op, post-op and recovery areas of hospital facilities. The intended user for this product is both Health Care Providers (HCPs: nurses, nurses aides, and medical doctors) and human patients. Lockout features may limit patient accessible controls.
Publicado el 9 de enero de 2013 · Inicio del retiro: 18 de diciembre de 2012
Resumen
FDA registra el retiro Z-0629-2013 de dispositivos médicos de Clase II, publicado el 9 de enero de 2013 e iniciado el 18 de diciembre de 2012. Empresa responsable: Stryker Medical Division of Stryker Corporation; Estado: Finalizado; Cantidad reportada: 1004; Lote: Model 3005S3EX, Serial Number 120415087 - 120515468,. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes.
Qué hacer
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Stryker Medical Division of Stryker Corporation
- Distribuidor
- Worldwide Distribution - Nationwide and Canada
- Lote
- Model 3005S3EX, Serial Number 120415087 - 120515468,
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAProCuity bed series, model number 3009, item number: 3009PX-LE-400.13 de diciembre de 2023
- FDAProform- support surface is used to assist in the prevention and treatment of all categories/stages of18 de agosto de 2021
- FDAPrime Fifth Wheel Stretcher-stretcher may be used for minor procedures and short-term stay (treatment and18 de agosto de 2021
- FDAThe Spirit Plus Bed, Model 5600 Spirit Plus /with high rail siderail option21 de noviembre de 2018
- FDAMedical Surgical Bed, Model 3002SEX Medical Surgical Bed The S3" MedSurg Bed, is intended to support and9 de enero de 2013
- FDADPM Central Station Mindray DS USA, Inc. 800 MacArthur Boulevard, Mahwah, NJ 07430 The Central Monitoring9 de enero de 2013
Preguntas sobre este retiro
¿Medical Surgical Bed, Model 3005S3EX Medical Surgical Bed The S3" MedSurg Bed, Model 3005 is intended to support and transport patients within the Med/Surg and Critical Care hospital environments. The bed is typically used in pre-op, post-op and recovery areas of hospital facilities. The intended user for this product is both Health Care Providers (HCPs: nurses, nurses aides, and medical doctors) and human patients. Lockout features may limit patient accessible controls. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0629-2013 el 9 de enero de 2013. Peligro reportado: Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit