Galaxy System (GAL-001)
Publicado el 25 de diciembre de 2024 · Inicio del retiro: 25 de octubre de 2024
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Noah Medical
- Distribuidor
- US: CA, MA, PA, NJ, ID, CT, MO, NC, MD, IL, TN, VA, OH, IN, WV, NY
- Lote
- UDI-DI: 00850048825048. Serial Numbers: CRB-1502, CRB-1503, CRB-1508, CRB-1513, CRB-1514, CRB-1515, CRB-1516, CRB-1518, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1528, CRB-1530, CRB-1532, CRB-1534, CRB-1535, CRB-1537, CRB-1538, CRB-1540, CRB-1541, CRB-1543, CRB-1554
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States