Centricity PACS System versions 3.X and higher and 4.X and higher Product Usage: The Centricity PACS Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. It is also intended for use as a clinical review workstation throughout the healthcare facility. the workstation interface provides the user with a means to display, manipulate, archive, print, and export images when connected with the Centricity PACS infrastructure.
Publicado el 22 de enero de 2014 · Inicio del retiro: 15 de septiembre de 2012
Resumen
FDA registra el retiro Z-0731-2014 de dispositivos médicos de Clase II, publicado el 22 de enero de 2014 e iniciado el 15 de septiembre de 2012. Empresa responsable: Ge Healthcare It; Estado: Finalizado; Cantidad reportada: 748 units; Lote: GE Centricity Software versions 3.X and higher and 4.X and higher. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
There is a potential loss of image(s) associated with the Centricity to Centricity (C2C) exam transfer module of Centricity PACS System versions 3.X and higher and 4.X and higher when transferring from the source to a destination server. When another process in the destination server attempts to access the same object or table, the transfer process of a particular image may be terminated. Once ter
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Ge Healthcare It
- Distribuidor
- Worldwide Distribution - US: Nationwide (AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WV, WY); FOREIGN: Australia, Austria, Belgium, Bermuda, China, Denmark, Egypt, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Kuwait, Malaysia, Malta, Netherlands, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.
- Lote
- GE Centricity Software versions 3.X and higher and 4.X and higher
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACentricity Enterprise Archive is a software product for receiving, archiving and sending of medical data.31 de diciembre de 2014
- FDAImagecast PACS is an image management system intended to be used by trained professionals, including but not22 de enero de 2014
- FDAPalacos® R+G Radioplaque Bone Cement 1 x 40 g Single (with Gentamicin), 40.8 g Methyl Acrylate Copolymer, 2022 de enero de 2014
- FDAVital Diagnostics Envoy Calcium Reagent For In-Vitro Diagnostic Use. PN 55285 For the quantitative measurement22 de enero de 2014
- FDAThe GE Centricity Web; Software Version 3.0x, is intended for use under the direct supervision of a licensed11 de diciembre de 2013
- FDACentricity PACS IW by GE Healthcare Dynamic Imaging Solutions is a device that receives medical images6 de noviembre de 2013
Preguntas sobre este retiro
¿Centricity PACS System versions 3.X and higher and 4.X and higher Product Usage: The Centricity PACS Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. It is also intended for use as a clinical review workstation throughout the healthcare facility. the workstation interface provides the user with a means to display, manipulate, archive, print, and export images when connected with the Centricity PACS infrastructure. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0731-2014 el 22 de enero de 2014. Peligro reportado: There is a potential loss of image(s) associated with the Centricity to Centricity (C2C) exam transfer module of Centricity PACS System versions 3.X and higher and 4.X and higher when transferring from the source to a destination server. When another process in the destination server attempts to access the same object or table, the transfer process of a particular image may be terminated. Once ter
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter