Imagecast PACS is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data and images are acquired through DICOM compliant imaging devices and modalities. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image and interpretations. Mammongraphic may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by the FDA.
Publicado el 22 de enero de 2014 · Inicio del retiro: 4 de marzo de 2011
Resumen
FDA registra el retiro Z-0773-2014 de dispositivos médicos de Clase II, publicado el 22 de enero de 2014 e iniciado el 4 de marzo de 2011. Empresa responsable: Ge Healthcare It; Estado: Finalizado; Cantidad reportada: 43 units; Lote: Imagecast PACS versions prior to 3.6.51.01. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
In Imagecast PACS versions prior to 3.6.51.01, the estimated magnification factor (0018,1114) was not accounted for. These versions of Imagecast PACS look for values to be present in the following tags to calculate a measurement from pixels to millimeters (mm): (0020,0030) Pixel Spacing and, (0018,1164) Imager Pixel Spacing. Because the estimated magnification factor is not used, distance meas
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Ge Healthcare It
- Distribuidor
- Nationwide Distribution: US including states of: AL, AZ, CT, FL, GA, IA, IN, LA, MA, MI, MN, MO, NC, NH, NJ, NY, OH, PA, TX, VA, WI, WV + District of Columbia.
- Lote
- Imagecast PACS versions prior to 3.6.51.01
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACentricity Enterprise Archive is a software product for receiving, archiving and sending of medical data.31 de diciembre de 2014
- FDACentricity PACS System versions 3.X and higher and 4.X and higher Product Usage: The Centricity PACS22 de enero de 2014
- FDABD Interlink threaded lock cannula This is a locking blunt plastic cannula specifically designed for IV22 de enero de 2014
- FDAEstrogen Receptor AB-11 (Clone ID5) Mouse Monoclonal Antibody 1 mk (2mg.ml) Manufactured by: Lab Vision22 de enero de 2014
- FDAThe GE Centricity Web; Software Version 3.0x, is intended for use under the direct supervision of a licensed11 de diciembre de 2013
- FDACentricity PACS IW by GE Healthcare Dynamic Imaging Solutions is a device that receives medical images6 de noviembre de 2013
Preguntas sobre este retiro
¿Imagecast PACS is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data and images are acquired through DICOM compliant imaging devices and modalities. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image and interpretations. Mammongraphic may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by the FDA. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0773-2014 el 22 de enero de 2014. Peligro reportado: In Imagecast PACS versions prior to 3.6.51.01, the estimated magnification factor (0018,1114) was not accounted for. These versions of Imagecast PACS look for values to be present in the following tags to calculate a measurement from pixels to millimeters (mm): (0020,0030) Pixel Spacing and, (0018,1164) Imager Pixel Spacing. Because the estimated magnification factor is not used, distance meas
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter