FDAFDA · Alimentos y medicamentos·N.º oficial Z-0807-2014

SOLAR 8000 AND TRANSPORT PRO WITH PATIENT DATA MODULE (Marketed As GE Healthcare CARESCAPE Patient Data Module) Product Usage: The Patient Data Module (PDM) is intended to provide uninterrupted acquisition of physiologic parameter data on adult, pediatric and neonatal patients during non-transport/bedside and transport patient care episodes. The Solar 8000i with Patient Data Module/Transport Pro with Patient Data Module is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional medical facility, such as hospital, clinic, surgical center or doctors office. It can be used in multiple areas such as operating room (OR), post anesthesia care unit (PACU), emergency department (ED), chest pain clinic, general intensive care unit (ICU), critical care unit, surgical intensive care unit (SICU), respiratory intensive care unit, coronary care unit (CCU), medical intensive care unit (MICU), pediatric intensive care unit (PICU), or neonatal intensive care unit (NICU). The Patient Data Module (PDM) is intended to provide uninterrupted acquisition of physiologic parameter data on adult, pediatric and neonatal patients during non-transport/bedside and transport patient care episodes. Physiological parameter data acquired by the PDM includes ECG, invasive pressure, non-invasive blood pressure, pulse oximetry, temperature, cardiac output nad respiration. This device acquires, processes and stores information for all aforementioned parameters and transmits this information to a transport or bedside central processing unit for viewing and alarm surveillance purposes. The Transport Pro patient monitor is intended for use as part of a transport monitoring system for intra-healthcare facility transport. When used with the Patient Data Module (PDM) or the TRAM acquisition module, this device is intended to provide uninterrupted monitoring of physiologic parameter data for adult, pediatric, and neonatal patients during transport from one area of the healthcare facility to another. Physiological parameter data includes ECG, invasive pressure, noninvasive blood pressure, pulse oximetry, temperature and respiration. Both the PDM and TRAM acquisition module acquire, process and store information for all aforementioned parameters. The Solar 8000i patient monitoring system is a multi-parameter physiological patient monitoring system intended for use on adult, pediatric and neonatal patients. It provides uninterrupted monitoring of physiological patient data. The Solar 8000i patient monitoring system is capable of monitoring and analyzing the following parameters for all patient populations: Electrocardiogram, invasive pressure, non-invasive blood pressure, pulse, temperature, cardiac output, respiration, pulse oximetry, venous oxygen saturation, transcutaneous pO2 and pCO2, CO2 and respiratory mechanics. The Solar 8000i patient monitoring system is capable of GE Healthcare Page 3 of 6 monitoring the following parameters for adult and pediatric patient populations: anesthetic agent concentrations, O2, impedance cardiography, electroencephalography and bispectral index. The Solar 8000i patient monitoring system interfaces with a variety of third-party peripheral medical devices that support serial and /or analog data outputs. Information from these devices can be displayed, trended and stored in the monitoring system. The solar 8000i patient monitoring system also provides physiological data over the Unity Network.

Publicado el 29 de enero de 2014 · Inicio del retiro: 5 de diciembre de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG cable connector on the DASH, PDM, and/or Tram Module. May result in reduced ECG, Respiration and arrhythmia detection performance.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
GE Healthcare, LLC
Distribuidor
Worldwide Distribution: US (nationwide) including DC and PR except SD, VT and WY; and Internationally to: ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, EGYPT, GERMANY, UNITED KINGDOM, IRELAND, IRAQ, ISRAEL, JAMAICA, JAPAN, JORDAN, KOREA, KUWAIT, LEBANON, LITHUANIA, MARITIUS, MALDOVA, NEW ZEALAND, NIGERIA, PAKISTAN, PANAMA, PARAGUAY, PERU, POLAND, QATAR, RUSSIA, SLOVAKIA, SOUTH AFRICA, SWEDEN, SWITZERLAND, TUNISIA, UNITED ARAB EMIRATES, VENEZUELA, CHILE, CANADA, and BERMUDA.
Lote
Serial Number SA308044521GR SA308186360GA SA310444859GA SA308186380GA SA312091241GA SA308491207GA SA308186401GA SA308186385GA SA308186374GA SA308155957GA SA308186391GA SA308186366GA SA308186410GA SA308094871GA SA308094818GA SA308186445GA SA308186381GA SA312479810GA SA308094940GA SA308094803GA SA308094872GA SA308084770GA SA310528992GA SA310506733GA SA310506696GA SA311153223GA SA311142931GA SA312080647GA SA312080894GA SA308186396GA SA308166121GA SA308186397GA SA308094807GA SA308186404GA SA311153169GA SA308186377GA SA308398894GA SA312253398GA SA308094809GA SA309263954GA SA310048616GA SA309527962GA SA310038386GA SA310048604GA SA310048613GA SA310028273GA SA310038462GA SA308317172GA SA311184814GA SA309365237GR SA311142946GA SA307332233GA SA307231341GA SA312069693GA SA309456714GA SA311163858GA SA312049242GA SA311317628GA SA307423146GA SA310259886GA SA309142960GA SA309334557GA SA307393020GA SA307453540GA SA307210827GA SA312018863GA SA307241567GA SA311163574GA SA312039090GA SA312080635GA SA309263938GA SA309304244GR SA309233517GA SA312039205GA SA312080692GA SA308347813GA SA308398654GA SA307504161GA SA312059549GA SA308491112GA SA312059550GA SA309021966GA SA310351543GA SA309193340GA SA308186334GA SA311101421GA SA310444513GA SA311518717GA SA312122048GA SA312029011GA SA309203388GA SA309344857GA SA312028978GA SA312253251GA SA312121695GA SA311508589GR SA312152342GA SA310517769GA SA310517770GA SA311297355GA SA311163926GA SA312029028GA SA308419235GA SA312069661GA SA311458372GA SA307382560GA SA311428153GA SA308511698GA SA308511500GA SA308368183GA SA309223482GA SA310361769GA SA312039183GA SA312079967GA SA308491092GA SA311458333GA SA310393014GA SA310507045GA SA310517560GA SA310331323GA SA311163587GA SA311184495GA SA307291912GA SA310393006GA SA311060085GA SA309112590GA SA310361759GA SA311194946GA SA309112586GA SA309112577GA SA309112584GA SA309112588GA SA309112583GA SA309112582GA SA309112591GA SA309112576GA SA309112569GA SA308166208GA SA308084740GA SA309112585GA SA309112578GA SA309112592GA SA309112572GA SA309112589GA SA309112587GA SA309112580GA SA309112579GA SA309122666GA SA309032010GA SA309042195GA SA308459639GA SA308449596GA SA311080599GA SA309092487GA SA309092489GA SA308145819GA SA308521831GA SA309032015GA SA309032018GA SA309032019GA SA309032020GA SA308459670GA SA309082457GA SA308501419GA SA308501431GA SA308501401GA SA308511444GA SA308481061GA SA309062308GA SA309062315GA SA309062313GA SA309062309GA SA309062310GA SA309062305GA SA309062300GA SA309082454GA SA309082456GA SA312469178GA SA309062269GA SA309021949GA SA309062277GA SA308511441GA SA308511440GA SA308511477GA SA309082443GA SA309092491GA SA309092492GA SA308327318GA SA308491248GA SA309062318GA SA309062296GA SA309062304GA SA309042201GA SA309042207GA SA308501411GA SA309062336GA SA309092493GA SA309082452GA SA311174002GA SA308317141GA SA309062298GA SA310518235GR SA308439427GA SA308398687GA SA308429319GA SA308429314GA SA308429310GA SA308439425GA SA308439442GA SA308439435GA SA308388629GA SA308419275GA SA308439462GA SA308429311GA SA308439431GA SA308429313GA SA308429316GA SA308419276GA SA308419277GA SA308439434GA SA308439426GA SA308439445GA SA308439461GA SA308439444GA SA308429320GA SA308439460GA SA308439433GA SA308429309GA SA308429317GA SA308429364GA SA308419278GA SA308439439GA SA311194970GA SA311184743GA SA311195201GA SA311184673GA SA309011857GA SA311287076GA SA310149349GA SA308459836GA SA310118991GA SA310068803GA SA308317131GA SA311428022GA SA308317113GA SA308266854GA SA308409143GA SA312049299GA SA308286919GA SA311488464GA SA309426162GA SA310507288GA SA310403433GA SA308388490GA SA307231395GA SA307453573GA SA307453582GA SA307504151GA SA307483662GA SA307483670GA SA307504093GA SA307504144GA SA307382578GA SA307504138GA SA307251801GA SA30750409
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States