Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0814-2024

TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal section, Sterile.

Publicado el 31 de enero de 2024 · Inicio del retiro: 18 de diciembre de 2023

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

When the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and instead reads and displays it as a different model. This results in some features specific to the catheter do not appear on the display and the catheter image may also appear inverted.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
St. Jude Medical, Atrial Fibrillation Division, Inc.
Distribuidor
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DC, DE, FL, GA, IA, IL, KS, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, and WV. The country of Canada.
Lote
Lot numbers 10073391, 10075618, 10080798, 10086492, 10087953, 10089624, 10092157, 10093066, and 10095363, Exp. 11/30/2024, GTIN 05415067034618.