Bone Screw, self-tapping The Trilogy shells are available in three configurations; uni, multi, and cluster-holed. The multi and cluster-holed shells allow for the use of screws to provide additional fixation and security, particularly in those case where acetabular bone stock is deficient. The Trilogy Acetabular System is indicated for either cemented or uncemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.

Publicado el 24 de diciembre de 2014 · Inicio del retiro: 21 de noviembre de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Analysis of returned complaint devices, product in inventory, and device manufacturing records has identified tool marks isolated to one specific machining mill to be the cause of the fractures. .

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Zimmer Manufacturing B.V.
Distribuidor
Worldwide Distribution - USA (nationwide) and Internationally to Canada Australia Brazil Chile China Colombia Japan Malaysia Singapore South Korea Taiwan Albania Austria Belgium Bulgaria Canary Islands Czech Republic Denmark Egypt Finland France Germany Greece Hungary Ireland Italy Netherlands Norway Poland Portugal Russia Saudi Arabia Serbia Slovakia South Africa Spain Sweden Switzerland Turkey and United Kingdom.
Lote
Item Number Lot Number 00-6250-065-35 62748089 00-6250-065-35 62754329 00-6250-065-35 62793494 00-6250-065-35 62793495 00-6250-065-35 62793501 00-6250-065-35 62793502 00-6250-065-35 62793503 00-6250-065-35 62813612 00-6250-065-35 62813613 00-6250-065-25 62784617 00-6250-065-25 62784618 2 00-6250-065-25 62784619 00-6250-065-25 62784621 00-6250-065-25 62784622 00-6250-065-25 62818701 00-6250-065-25 62818702 00-6250-065-25 62818703 00-6250-065-25 62818707 00-6250-065-25 62818709 00-6250-065-25 62818710 00-6250-065-25 62825611 00-6250-065-25 62836984
Clasificación de riesgo
Class II
Área (FDA)
device
País
US