Gendex GXPS-500 SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)

Publicado el 29 de enero de 2020 · Inicio del retiro: 7 de enero de 2020

Resumen

FDA registra el retiro Z-0840-2020 de dispositivos médicos de Clase II, publicado el 29 de enero de 2020 e iniciado el 7 de enero de 2020. Empresa responsable: Palodex Group Oy; Estado: Finalizado; Cantidad reportada: 101; Lote: Serial Number: GJ1209488, GJ1209489, GJ1209491, GJ1209492, GJ1209493, GJ1209494, GJ1209495, GJ1209496, GJ1209498, GJ1209499, GJ1209504, GJ1209505, GJ1209506, GJ1209507, GJ1209509, GJ1209510, GJ1209511, GJ1209513, GJ1209514, GJ1209516, GJ1211287, GJ1211288, GJ1211289, GJ1211290, GJ1211291, GJ1211292, GJ1211293, GJ1211294, GJ1211295, GJ1211296, GJ1211297, GJ1211298, GJ1211299, GJ1211300, GJ1211301, GJ1211302, GJ1211303, GJ1211304, GJ1211305, GJ1211307, GJ1311619, GJ1311621, GJ1311622, GJ1311623, GJ1311624, GJ1311625, GJ1311626, GJ1311627, GJ1311628, GJ1311629, GJ1311630, GJ1311631, GJ1311632, GJ1311633, GJ1311634, GJ1311635, GJ1311636, GJ1311637, GJ1311638, GJ1311639, GJ1311640, GJ1311641, GJ1311642, GJ1311643, GJ1311644, GJ1311645, GJ1311646, GJ1311647, GJ1311648, GJ1311650, GJ1311651, GJ1311653, GJ1311656, GJ1311658, GJ1311659, GJ1311660, GJ1311661, GJ1311662, GJ1311663, GJ1311664, GJ1311665, GJ1311666, GJ1311669, GJ1311670, GJ1311671, GJ1311672, GJ1311673, GJ1311674, GJ1311675, GJ1311676, GJ1311677, GJ1311678, GJ1311679, GJ1311680, GJ1311681, GJ1311682, GJ1311683, GJ1311684, GJ1311685, GJ1311686, GJ1311722 PSU's having nine-digit serial numbers beginning with four digits as follows: 1143XXXXX 1221XXXXX 1224XXXXX and 1229XXXXX. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Palodex Group Oy
Distribuidor
US: WI and PA OUS: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KOREA, REPUBLIC OF, KUWAIT, MALAYSIA, MALTA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PAKISTAN, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, REPUBLIC OF CHINA, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM
Lote
Serial Number: GJ1209488, GJ1209489, GJ1209491, GJ1209492, GJ1209493, GJ1209494, GJ1209495, GJ1209496, GJ1209498, GJ1209499, GJ1209504, GJ1209505, GJ1209506, GJ1209507, GJ1209509, GJ1209510, GJ1209511, GJ1209513, GJ1209514, GJ1209516, GJ1211287, GJ1211288, GJ1211289, GJ1211290, GJ1211291, GJ1211292, GJ1211293, GJ1211294, GJ1211295, GJ1211296, GJ1211297, GJ1211298, GJ1211299, GJ1211300, GJ1211301, GJ1211302, GJ1211303, GJ1211304, GJ1211305, GJ1211307, GJ1311619, GJ1311621, GJ1311622, GJ1311623, GJ1311624, GJ1311625, GJ1311626, GJ1311627, GJ1311628, GJ1311629, GJ1311630, GJ1311631, GJ1311632, GJ1311633, GJ1311634, GJ1311635, GJ1311636, GJ1311637, GJ1311638, GJ1311639, GJ1311640, GJ1311641, GJ1311642, GJ1311643, GJ1311644, GJ1311645, GJ1311646, GJ1311647, GJ1311648, GJ1311650, GJ1311651, GJ1311653, GJ1311656, GJ1311658, GJ1311659, GJ1311660, GJ1311661, GJ1311662, GJ1311663, GJ1311664, GJ1311665, GJ1311666, GJ1311669, GJ1311670, GJ1311671, GJ1311672, GJ1311673, GJ1311674, GJ1311675, GJ1311676, GJ1311677, GJ1311678, GJ1311679, GJ1311680, GJ1311681, GJ1311682, GJ1311683, GJ1311684, GJ1311685, GJ1311686, GJ1311722 PSU's having nine-digit serial numbers beginning with four digits as follows: 1143XXXXX 1221XXXXX 1224XXXXX and 1229XXXXX
Clasificación de riesgo
Class II
Área (FDA)
device
País
Finland

Retiros relacionados

Preguntas sobre este retiro

¿Gendex GXPS-500 SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER) fue retirado del mercado?

Sí. FDA publicó el retiro Z-0840-2020 el 29 de enero de 2020. Peligro reportado: Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter