GE Healthcare Innova IGS 630. Biplane Cardiovascular and Interventional Imaging System. Product Usage: The angiographic X-ray systems are indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image-guided surgical procedures. The OR table is suitable for interventional and surgical procedures. The Cathiab Frontiers solutions are indicated for use in conjunction with single plane and biplane GE angiographic X-ray systems and imaging / data medical devices used in interventional and surgical cathlab environments and cleared for commercial distribution. The Cathiab Frontiers solutions are integrated GE angiographic X-ray and imaging / data medical devices that simplify the end-to-end clinical workflow in the cathlab by implementing: 1) Communication protocols for exchanging and automatically synchronizing patient, exam, system, and image information between the angiographic X-ray systems and the imaging / data medical devices. 2) Communication protocols for the control of imaging / data medical device functions from the angiographic X-ray systems user interface. 3) Interfaces for displaying the imaging / data medical device output on the monitor display solutions of the GE angiographic X-ray systems.
Publicado el 5 de febrero de 2014 · Inicio del retiro: 15 de noviembre de 2013
Resumen
FDA registra el retiro Z-0853-2014 de dispositivos médicos de Clase II, publicado el 5 de febrero de 2014 e iniciado el 15 de noviembre de 2013. Empresa responsable: GE Healthcare, LLC; Estado: Finalizado; Cantidad reportada: 28 (12 US, 16 OUS).; Lote: Serial # System ID 00000616724BU1 956389IGS630 00000618156BU4 MCMIGS630 FMI12200DUMMY2 702731ANGIO2 00000619372BU6 732828IGS30 00000618158BU0 210705CATH1 00000622076BU9 661327IGS630 00000621314BU4 904202WCL2 00000628656BU1 281MHWCL1 00000622220BU2 303839CL4 00000625357BU9 281420LAB3BP 00000626085BU5 817BGIGS630 00000628654BU6 215576IGS4 FMI12200DUMMY3 CS1340VA01 00000614683BU1 M4152252 00000618160BU6 YV1900 00000617143BU3 YV1901 00000618157BU2 YV1902 00000620495BU2 YV1903 00000624842BU1 YV1904 00000627935BU0 YV1905 00000630300BU2 YV1906 00000617142BU5 YV1950 FMI12200DUMMY4 YV1907 00000622672BU4 850060746 00000626239BU8 850060749 FMI12200DUMMY1 SA2308VA01 00000622221BU0 ZA2239VA01 00000614682BU3 34006VAS02. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
GE Healthcare has recently become aware of a potential safety issue with respect to IGS 630 Imaging Systems. While performing fluoroscopy on a biplane fluoroscopy unit, there is a potential for loss of the x-ray imaging function when the user changes the size of the lateral FOV (Field of View) and releases the biplane footswitch pedal simultaneously. This could lead to a system lock up requiring
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- GE Healthcare, LLC
- Distribuidor
- Worldwide Distribution - US Nationwide in the states of CA, CO, FL, NV, NJ, PA, TX and the countries of CZECH REPUBLIC, FRANCE, JAPAN, REPUBLIC OF KOREA, SAUDI ARABIA, and TURKEY.
- Lote
- Serial # System ID 00000616724BU1 956389IGS630 00000618156BU4 MCMIGS630 FMI12200DUMMY2 702731ANGIO2 00000619372BU6 732828IGS30 00000618158BU0 210705CATH1 00000622076BU9 661327IGS630 00000621314BU4 904202WCL2 00000628656BU1 281MHWCL1 00000622220BU2 303839CL4 00000625357BU9 281420LAB3BP 00000626085BU5 817BGIGS630 00000628654BU6 215576IGS4 FMI12200DUMMY3 CS1340VA01 00000614683BU1 M4152252 00000618160BU6 YV1900 00000617143BU3 YV1901 00000618157BU2 YV1902 00000620495BU2 YV1903 00000624842BU1 YV1904 00000627935BU0 YV1905 00000630300BU2 YV1906 00000617142BU5 YV1950 FMI12200DUMMY4 YV1907 00000622672BU4 850060746 00000626239BU8 850060749 FMI12200DUMMY1 SA2308VA01 00000622221BU0 ZA2239VA01 00000614682BU3 34006VAS02
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACentricity PACS Software Version 7.0 SP0.0.4.73 de agosto de 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.27 de julio de 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images and13 de julio de 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 20196 de julio de 2022
- FDAThe ACUTE Innovations Modular RibLoc System (RibLoc U plus) is comprised titanium plates and screws that are5 de febrero de 2014
- FDAGE Healthcare CARESCAPE Monitor B650 and CARESCAPE Monitor B850. The CARESCAPE Monitor B650 is a5 de febrero de 2014
Preguntas sobre este retiro
¿GE Healthcare Innova IGS 630. Biplane Cardiovascular and Interventional Imaging System. Product Usage: The angiographic X-ray systems are indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image-guided surgical procedures. The OR table is suitable for interventional and surgical procedures. The Cathiab Frontiers solutions are indicated for use in conjunction with single plane and biplane GE angiographic X-ray systems and imaging / data medical devices used in interventional and surgical cathlab environments and cleared for commercial distribution. The Cathiab Frontiers solutions are integrated GE angiographic X-ray and imaging / data medical devices that simplify the end-to-end clinical workflow in the cathlab by implementing: 1) Communication protocols for exchanging and automatically synchronizing patient, exam, system, and image information between the angiographic X-ray systems and the imaging / data medical devices. 2) Communication protocols for the control of imaging / data medical device functions from the angiographic X-ray systems user interface. 3) Interfaces for displaying the imaging / data medical device output on the monitor display solutions of the GE angiographic X-ray systems. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0853-2014 el 5 de febrero de 2014. Peligro reportado: GE Healthcare has recently become aware of a potential safety issue with respect to IGS 630 Imaging Systems. While performing fluoroscopy on a biplane fluoroscopy unit, there is a potential for loss of the x-ray imaging function when the user changes the size of the lateral FOV (Field of View) and releases the biplane footswitch pedal simultaneously. This could lead to a system lock up requiring
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter