EkoSonic Endovascular System, Catalog Number 500-56112. It consists of three main components, a single use EkoSonic Endovascular Device which is comprised of an Intelligent Drug Delivery Catheter (IDDC) and removable MicroSonic Device (MSD), and a reusable EKOS EkoSonic Control System. During use, fluids are delivered through the side-holes of the IDDC. The MSD is placed through the central lumen of the IDDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS EkoSonic Control Unit provides power to the system and the user interface for operator control. A reusable, non-sterile Connector Interface Cable connects the EKOS EkoSonic Control Unit to the EkoSonic Endovascular Device.
Publicado el 2 de marzo de 2016 · Inicio del retiro: 21 de enero de 2016
Resumen
FDA registra el retiro Z-0864-2016 de dispositivos médicos de Clase II, publicado el 2 de marzo de 2016 e iniciado el 21 de enero de 2016. Empresa responsable: EKOS Corporation; Estado: Finalizado; Cantidad reportada: 10 units; Lote: Lot Number 150910004. Expiration date is August 15, 2018. Serial numbers: 001, 002, 003, 004, 005, 006, 007, 008, 009, 010. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The EkoSonic kit, including the individually labeled MicroSonic Device (MSD) and Intelligent Drug Delivery Catheter (IDDC) are labeled with a 135 cm working length, and programmed with a 135 cm family code, however the MSD actual working length is 106 cm.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- EKOS Corporation
- Distribuidor
- Distributed in the states of AZ, CA, MA, MI, OH, and VA.
- Lote
- Lot Number 150910004. Expiration date is August 15, 2018. Serial numbers: 001, 002, 003, 004, 005, 006, 007, 008, 009, 010
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAThe Symbiq infusion pump is a device used in a health care facility to pump fluids into a patient in a2 de marzo de 2016
- FDAFlexor Check-Flo Introducer, Catheter introducer, Catalog #: KCFW-6.0-35-70-RB-HFANL0-HC Intended for2 de marzo de 2016
- FDASyngo Plaza Picture archiving and communication system. Software only. PACS intended to display, process,2 de marzo de 2016
- FDAStryker, Gamma3 K-Wire with Washer, 13x3.2x500 mm Product Usage: Intended for the temporary stabilization of2 de marzo de 2016
- FDAThe EkoSonic Endovascular System employs high frequency (2-2.5 MHz), low power (0.585 watt average power per13 de noviembre de 2013
- FDAEkoSonic MACH4 Endovascular Device (Catalog #500-55106, 500-55112; 500-55118; 500-55124; 500-55130; 500-55140;25 de julio de 2012
Preguntas sobre este retiro
¿EkoSonic Endovascular System, Catalog Number 500-56112. It consists of three main components, a single use EkoSonic Endovascular Device which is comprised of an Intelligent Drug Delivery Catheter (IDDC) and removable MicroSonic Device (MSD), and a reusable EKOS EkoSonic Control System. During use, fluids are delivered through the side-holes of the IDDC. The MSD is placed through the central lumen of the IDDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS EkoSonic Control Unit provides power to the system and the user interface for operator control. A reusable, non-sterile Connector Interface Cable connects the EKOS EkoSonic Control Unit to the EkoSonic Endovascular Device. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0864-2016 el 2 de marzo de 2016. Peligro reportado: The EkoSonic kit, including the individually labeled MicroSonic Device (MSD) and Intelligent Drug Delivery Catheter (IDDC) are labeled with a 135 cm working length, and programmed with a 135 cm family code, however the MSD actual working length is 106 cm.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter