EkoSonic Endovascular System, Catalog Number 500-56112. It consists of three main components, a single use EkoSonic Endovascular Device which is comprised of an Intelligent Drug Delivery Catheter (IDDC) and removable MicroSonic Device (MSD), and a reusable EKOS EkoSonic Control System. During use, fluids are delivered through the side-holes of the IDDC. The MSD is placed through the central lumen of the IDDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS EkoSonic Control Unit provides power to the system and the user interface for operator control. A reusable, non-sterile Connector Interface Cable connects the EKOS EkoSonic Control Unit to the EkoSonic Endovascular Device.

Publicado el 2 de marzo de 2016 · Inicio del retiro: 21 de enero de 2016

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The EkoSonic kit, including the individually labeled MicroSonic Device (MSD) and Intelligent Drug Delivery Catheter (IDDC) are labeled with a 135 cm working length, and programmed with a 135 cm family code, however the MSD actual working length is 106 cm.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
EKOS Corporation
Distribuidor
Distributed in the states of AZ, CA, MA, MI, OH, and VA.
Lote
Lot Number 150910004. Expiration date is August 15, 2018. Serial numbers: 001, 002, 003, 004, 005, 006, 007, 008, 009, 010
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States