Trilogy¿ Fiber Metal Acetabular Shell Size 54 mm OD The Trilogy Acetabular System is indicated for either cemented or uncemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.
Publicado el 31 de diciembre de 2014 · Inicio del retiro: 4 de diciembre de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Zimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigation found the polar-hole threads to be present but non-conforming to specifications, in that the minor diameter was undersized. Investigation also indicates that the issue is isolated and affects only lot 62714305, which
Qué hacer
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Detalles del retiro
- Fabricante
- Zimmer Manufacturing B.V.
- Distribuidor
- Worldwide Distribution: US (nationwide) and countries of: France, Germany, Iran, Romania, Serbia, Singapore, Taiwan, and United Kingdom.
- Lote
- Item No. 00-6200-054-20; Lot 62714305
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States