MitraClip Clip Delivery System, product number MSK0101. The MitraClip System contains the Clip Delivery System, product number CDS0201, GTIN 08717648195914 and the Steerable Guide Catheter, product number SGC0101, GTIN 08717648195921. The MitraClip Clip Delivery System (CDS) consists of three major components: 1) the Delivery Catheter 2) the Steerable Sleeve, and 3) the MitraClip device. The implantable MitraClip device is located at the distal end of the CDS. The CDS is used to advance and manipulate the implantable MitraClip device for proper positioning and placement on the mitral valve leaflets. The Delivery Catheter, which is part of the CDS, is designed to deliver and deploy the MitraClip. The Delivery Catheter extends through the thru-lumen of the Steerable Sleeve.
Publicado el 16 de marzo de 2016 · Inicio del retiro: 4 de febrero de 2016
Resumen
FDA registra el retiro Z-0911-2016 de dispositivos médicos de Clase I, publicado el 16 de marzo de 2016 e iniciado el 4 de febrero de 2016. Empresa responsable: Abbott Vascular; Estado: Finalizado; Cantidad reportada: 7,600 units total (2,300 units in US); Lote: Lot number 50811U1 50811U2 50812U1 50813U1 50814U1 50826U1 50826U2 50827U1 50908U1 50908U2 50909U1 50910U1 50910U2 50911U1 50924U1 50925U1 50928U1 50929U1 51010U1 51012U1 51012U2 51013U1 51014U1 51014U2 51026U1 51027U3 51028U1 51028U2 51029U1 51030U1 51105U1 51106U1 51109U1 51109U2 51110U1 51110U2 51117U1 51203U1 51204U1 51205U1 51207U1. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Abbott Vascular has recently received reports of cases on Clip Delivery System devices that contain the One-Way Actuator Knob where a user attempted implanting a MitraClip, but the Clip could not be detached from the delivery system due to a mandrel fracture.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Abbott Vascular
- Distribuidor
- Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Argentina Luxembourg Australia Malaysia Austria Mexico Belgium Netherlands Brazil New Zealand Canada Norway Colombia Philippines Corrientes Poland Czech Republic Portugal Denmark Egypt Romania Finland Saudi Arabia France Slovenia Germany Spain Greece Sweden Hong Kong Switzerland Hungary Thailand Indonesia Turkey Israel United Arab Emirates Italy United Kingdom Kuwait and Vietnam.
- UPC
- 08717648195914
- Lote
- Lot number 50811U1 50811U2 50812U1 50813U1 50814U1 50826U1 50826U2 50827U1 50908U1 50908U2 50909U1 50910U1 50910U2 50911U1 50924U1 50925U1 50928U1 50929U1 51010U1 51012U1 51012U2 51013U1 51014U1 51014U2 51026U1 51027U3 51028U1 51028U2 51029U1 51030U1 51105U1 51106U1 51109U1 51109U2 51110U1 51110U2 51117U1 51203U1 51204U1 51205U1 51207U1
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAPlus 30 Priority Pack w/.096 RHV - REF 1000183 WPL2122277-01 (20-03-01) Plus 30 Priority Pack w/.115 RHV - REF20 de abril de 2022
- FDANC TREK Coronary Dilatation Catheter, REF 1012453-20 Rx Only, NC TREK RX 4.50 X 8MM BDC The NC TREK RX26 de febrero de 2020
- FDANC TRAVELER Coronary Dilatation Catheter, REF 1013157-15, Rx Only,NC Traveler RX 4.0 X 15MM The NC Traveler RX26 de febrero de 2020
- FDANC TREK Coronary Dilatation Catheter, REF 1012453-20 Rx Only, NC TREK RX 4.50 X 8MM BDC The NC TREK RX26 de febrero de 2020
- FDAUrban Oven Artisan-Baked Short Bread Cookies Lemon Pistachio All Natural NET WT. 5 oz UPC Code: 7 18122 0637416 de marzo de 2016
- FDAM/L Taper Hip Prosthesis (Item: 00-7711-012-20, Lot: 6293709316 de marzo de 2016
Preguntas sobre este retiro
¿MitraClip Clip Delivery System, product number MSK0101. The MitraClip System contains the Clip Delivery System, product number CDS0201, GTIN 08717648195914 and the Steerable Guide Catheter, product number SGC0101, GTIN 08717648195921. The MitraClip Clip Delivery System (CDS) consists of three major components: 1) the Delivery Catheter 2) the Steerable Sleeve, and 3) the MitraClip device. The implantable MitraClip device is located at the distal end of the CDS. The CDS is used to advance and manipulate the implantable MitraClip device for proper positioning and placement on the mitral valve leaflets. The Delivery Catheter, which is part of the CDS, is designed to deliver and deploy the MitraClip. The Delivery Catheter extends through the thru-lumen of the Steerable Sleeve. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0911-2016 el 16 de marzo de 2016. Peligro reportado: Abbott Vascular has recently received reports of cases on Clip Delivery System devices that contain the One-Way Actuator Knob where a user attempted implanting a MitraClip, but the Clip could not be detached from the delivery system due to a mandrel fracture.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter
¿Cómo sé si mi unidad está incluida en el retiro?
Compara los datos de tu producto con los del aviso. UPC: 08717648195914.