Trumpf Mount Clamp Assembly (table adapter component), part #118704-000, a component of the IMRIS HFD100 head fixation device, models #119629-000 and #119630-000, which are used with ORT400 operating room tables.

Publicado el 20 de abril de 2022 · Inicio del retiro: 17 de febrero de 2022

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Reports have been received that after normal or routine lateral forces are applied to the head fixation device, the split block clamp may deflect (rotationally around the axis) and not return to within acceptable distance from its prior position after the force is removed. In order to reduce deflection at the table mount location, users have tightened the hand knob screw to a degree that it is then difficult to untighten it later in the clinical workflow. Rotational movement at the table attachment point could result in HFD movement going unnoticed, which could result in a situation where surgical or medical intervention is necessary, a situation where navigation needs to be reestablished, or a situation resulting in extended time under anesthesia for the patient. It may also be difficult to loosen the mount clamp component after use, potentially leading to ergonomic injury or a workaround during equipment teardown.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Deerfield Imaging, Inc.
Distribuidor
Worldwide distribution - US Nationwide distribution in the states of AL, FL, GA, MA, MN, NC, NY, OH, and TX. The countries of Sweden and Switzerland.
Lote
Model #119629-000, UDI 00857534006592; and model #119630-000, UDI 00857534006615. Serial numbers: 10008164, 10007550, 10007555, 10008864, 10008160, 10008161, 10008859, 10008860, 10010668, 10010669, 10010673, 10010674, 10010672, 10007556, 10007557, 10008862, 10008861, 10008158, 10008159, 10008863, 10008165, 10007558, 10008865, and 10010675.
Clasificación de riesgo
Class II
Área (FDA)
device