LOCATOR® Implant Attachment System. The LOCATOR® Implant Attachment System is a universal hinge, resilient attachment for endosseous implants. The LOCATOR® Implant Attachment System is designed for use with overdentures or partial dentures, retained in whole or in part, by endosseous implants in the mandible or maxilla. FDA 510(k): K994257, K072878
Publicado el 12 de febrero de 2014 · Inicio del retiro: 16 de enero de 2014
Resumen
FDA registra el retiro Z-0935-2014 de dispositivos médicos de Clase II, publicado el 12 de febrero de 2014 e iniciado el 16 de enero de 2014. Empresa responsable: Zest Anchors LLC; Estado: Finalizado; Cantidad reportada: 3,130; Lote: Part Numbers: 8875-SPI, 8875-SPI-TMS, 8876-SPI, 8876-SPI-TMS, 08877-SPI, 08877-SPI-TMS, 08878-SPI, 08878-SPI-TMS, 08879-SPI, 08880-SPI; Lot Numbers: 11052, 15306, 20665, 21351, 12025,12259, 22440, I0K7X, I0K7Y, 11963,15307,17676,18934, 25294, 25567, I0JQ7, 12258,12941,15000, 19735, 20666, I0K7Z, I0KY7, 12703, 14568, 19832, 22170, I0LCG, 11093, 13364, 14151, 15624, 20043, 25026, 13530, I0LE4, 10055, 14152, 15001, 25027, 10892, 10893. Since the LOCATOR® Implant Attachment System (including the restorative components) is not sold by the Zest Anchors as a sterilized product, it does not have a designated shelf-life. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Zest Anchors, LLC is voluntarily recalling specific lots of LOCATOR® Implant Attachment Systems because it is currently provided with a label that indicates that the final torque applied to the abutment should be 30 N-cm. It was recently noted by Zest that in order to provide an optimal performance by increasing the factor of safety between the failure load of the abutment screw and the recommend
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Zest Anchors LLC
- Distribuidor
- Distributed in Ohio and Switzerland.
- Lote
- Part Numbers: 8875-SPI, 8875-SPI-TMS, 8876-SPI, 8876-SPI-TMS, 08877-SPI, 08877-SPI-TMS, 08878-SPI, 08878-SPI-TMS, 08879-SPI, 08880-SPI; Lot Numbers: 11052, 15306, 20665, 21351, 12025,12259, 22440, I0K7X, I0K7Y, 11963,15307,17676,18934, 25294, 25567, I0JQ7, 12258,12941,15000, 19735, 20666, I0K7Z, I0KY7, 12703, 14568, 19832, 22170, I0LCG, 11093, 13364, 14151, 15624, 20043, 25026, 13530, I0LE4, 10055, 14152, 15001, 25027, 10892, 10893. Since the LOCATOR® Implant Attachment System (including the restorative components) is not sold by the Zest Anchors as a sterilized product, it does not have a designated shelf-life.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAOxcarbazepine Tablets, 300 mg, 100 tablet bottles, Rx Only. Distributed by: Caraco Pharmaceutical19 de febrero de 2014
- FDA3.5mm Locking Cortical Bone Screw, T15 Hexalobe, Self Tapping, 20mm Part No. 00-0903-2620 OrthoPediatrics12 de febrero de 2014
- FDAThe Symbionic Leg is an integrated prosthetic leg for transfemoral and knee disarticulation amputees. The12 de febrero de 2014
- FDAThe Gaymar Medi-Therm device provides a means of regulating patient temperature by supplying12 de febrero de 2014
- FDAHemochron Jr. Citrate APTT Cuvette 510(k) K014008 A unitized microcoagulation test intended to be used in12 de febrero de 2014
- FDADiagnostic imaging systems (GEMINI LXL Model #882412, TruFlight Select Model #882438, GEMINI TF PET/CT16 Model12 de febrero de 2014
Preguntas sobre este retiro
¿LOCATOR® Implant Attachment System. The LOCATOR® Implant Attachment System is a universal hinge, resilient attachment for endosseous implants. The LOCATOR® Implant Attachment System is designed for use with overdentures or partial dentures, retained in whole or in part, by endosseous implants in the mandible or maxilla. FDA 510(k): K994257, K072878 fue retirado del mercado?
Sí. FDA publicó el retiro Z-0935-2014 el 12 de febrero de 2014. Peligro reportado: Zest Anchors, LLC is voluntarily recalling specific lots of LOCATOR® Implant Attachment Systems because it is currently provided with a label that indicates that the final torque applied to the abutment should be 30 N-cm. It was recently noted by Zest that in order to provide an optimal performance by increasing the factor of safety between the failure load of the abutment screw and the recommend
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter