Algerbrush-II Product Usage: Medical - For surgeons to use in ophthalmic surgical procedures to remove foreign bodies from the eye of the patient

Publicado el 18 de enero de 2017 · Inicio del retiro: 12 de octubre de 2016

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

According to firm ( Bausch & Lomb, Inc. ) on June 17, 2016 a packaging operator notified his supervisor he had not placed the operating instruction into the last repackaging of SKU 0816 C (Pterygium Burr Medium Diamond). Further investigation found that repackaged Algerbrush-II and its accessories from April 30, 2012 to present, contained out of date operating instruction. Therefore Bausch & Lomb is proposing a Field Correction to provide customer purchasing product from April 30, 2012 to June 20, 2016 with the current operating instruction.

Qué hacer

Voluntary: Firm initiated · E-Mail

Detalles del retiro

Fabricante
Bausch & Lomb, Inc.
Distribuidor
Worldwide Distribution - US Nationwide
Lote
LOT #(s) AND EXPIRATION DATES, SERIAL #(s), ORDER #, or OTHER IDENTIFICATION #(s): SKU E0815 A, SKU E0815 B, SKU E0815 0.5, SKU E0815 1.0 and SKU E0816 C
Clasificación de riesgo
Class II