Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0990-2013

VITEK 2 XL computer system when equipped with the following Hewlett-Packard PC models: (1) rp5700, Cat. #W0452; (2) rp5700 (refurbished), Cat. #W0452R; (3) rp5800 XPE, Cat. 413642; (4) rp5800 WES7, Cat. #413862; (5) dc7100 8-port, Cat. #W0441; (6) dc7100 8-port (refurbished), Cat. #W0441R; (7) dc7700 8-port, Cat. #W0447; (8) dc7700 8-port (refurbished), Cat. #W0447R; (9) dc7800, Cat. #W0449; and (10) dc7800 (refurbished), Cat. #W0449R. VITEK 2 is an automated system designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.

Publicado el 3 de abril de 2013 · Inicio del retiro: 18 de febrero de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.

Qué hacer

Voluntary: Firm initiated · E-Mail

Detalles del retiro

Fabricante
Biomerieux Inc
Distribuidor
Worldwide Distribution-USA (nationwide) and the countries of Algeria, Angola, Argentina, Australia, Austria, Azerbaidjian, Bahrain, Belarus, Belgium, Benin, Bolivia, Bosnia, Brazil, Bulgaria, Burkina Faso, Camaroun, Canada, Chad, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Democratic Republic of Congo, Demark, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Ethiopia, Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Guinea, Herzegovina, Honduras, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mali, Mexico, Moldavia, Mongolia, Morocco, Myanmar, Netherlands, Netherlands, Nicaragua, Niger Niamey, Nigeria, Norway, Oman, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, and Yemen.
Lote
Manufactured from September, 2004, to present.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States