Air-Q3 AirQ3 Standard ILA with Manual Cuff Inflation and GA Size REF Number 0.0 50005 0.5 50055 1.0 50105 1.5 50155 2.0 50205 3.0 50305 4.0 50405
Publicado el 1 de febrero de 2023 · Inicio del retiro: 22 de diciembre de 2022
Resumen
FDA registra el retiro Z-1019-2023 de dispositivos médicos de Clase II, publicado el 1 de febrero de 2023 e iniciado el 22 de diciembre de 2022. Empresa responsable: SunMed Holdings, LLC; Estado: En curso; Cantidad reportada: 43,290 (Eaches); Lote: ALL LOTS Size REF UDI/DI 0.0 50005 Each: 00814954020188 Box: 10814954020185 Case: 20814954020182 0.5 50055 Each: 00814954020195 Box: 10814954020192 Case: 20814954020199 1.0 50105 Each: 00814954020201 Box: 10814954021007 Case: 20814954020205 1.5 50155 Each: 00814954020218 Box: 10814954021014 Case: 20814954020212 2.0 50205 Each: 00814954020225 Box: 10814954021021 Case: 20814954020229 3.0 50305 Each: 00814954020232 Box: 10814954021038 Case: 20814954020236 4.0 50405 Each: 00814954020249 Box: 10814954021045 Case: 20814954020243. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
There is an orogastric (OG) tube size discrepancy between the labeling in the IFU and the printed text on the device. If the clinician is unable to pass a gastric catheter through one of the channels, the clinician may use a smaller-diameter gastric catheter.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- SunMed Holdings, LLC
- Distribuidor
- US Nationwide distribution in the states of AL, CA, FL, MI, OH, TN, TX, WA.
- Lote
- ALL LOTS Size REF UDI/DI 0.0 50005 Each: 00814954020188 Box: 10814954020185 Case: 20814954020182 0.5 50055 Each: 00814954020195 Box: 10814954020192 Case: 20814954020199 1.0 50105 Each: 00814954020201 Box: 10814954021007 Case: 20814954020205 1.5 50155 Each: 00814954020218 Box: 10814954021014 Case: 20814954020212 2.0 50205 Each: 00814954020225 Box: 10814954021021 Case: 20814954020229 3.0 50305 Each: 00814954020232 Box: 10814954021038 Case: 20814954020236 4.0 50405 Each: 00814954020249 Box: 10814954021045 Case: 20814954020243
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAMedline manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Adult29 de noviembre de 2023
- FDAVentlab, LLC AirFlow manual resuscitator devices with integrated manometer, multiple accessory configurations29 de noviembre de 2023
- FDAEasy Care first aid Burn Cream (benzalkonium chloride 0.13% and lidocaine hydrochloride 0.5%), 0.9 g (0.03 oz)8 de febrero de 2023
- FDAAdventure Marine 150, First Aid Kit, containing Easy Care first aid After Burn Cream 0.9 g packets, Product #8 de febrero de 2023
- FDARevogene SARS-CoV-2_IVD real-time RT-PCR test intended for the qualitative detection of RNA from SARSCoV-21 de febrero de 2023
- FDAChef Select 2 Herb Baguettes Filled With Herb Butter 175 g Barcode: 40564894475661 de febrero de 2023
Preguntas sobre este retiro
¿Air-Q3 AirQ3 Standard ILA with Manual Cuff Inflation and GA Size REF Number 0.0 50005 0.5 50055 1.0 50105 1.5 50155 2.0 50205 3.0 50305 4.0 50405 fue retirado del mercado?
Sí. FDA publicó el retiro Z-1019-2023 el 1 de febrero de 2023. Peligro reportado: There is an orogastric (OG) tube size discrepancy between the labeling in the IFU and the printed text on the device. If the clinician is unable to pass a gastric catheter through one of the channels, the clinician may use a smaller-diameter gastric catheter.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter