Medline manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Adult Resuscitator REF #s: CPRM1116FPM, CPRM1116FPM4, CPRM1116M, CPRM1116MC, CPRM1116PM, CPRM1116PMT, CPRM1126FM, CPRM1126FPM, CPRM1126FPWM, CPRM1126M; 2. Infant Resuscitator REF #s: CPRM3312FM, CPRM3312FM123, CPRM3312FPM, CPRM3312FPM3, CPRM3312FPMC, CPRM3312M, CPRM3312M3, CPRM3312P3M3, CPRM3312PM, CPRM3312PM2, CPRM3312PMC2, CPRM3312PMTD1, CPRM3312PMX, CPRM3312PWM, CPRM3322FM, CPRM3322M, CPRM3322PM; 3. Pediatric Resuscitator REF #s: CPRM2216FM, CPRM2216FM3, CPRM2216FPM, CPRM2216FPM3, CPRM2216FPMC, CPRM2216M, CPRM2216P3M3, CPRM2216PCM, CPRM2216PM, CPRM2216PM6, CPRM2226FM, CPRM2226FM3, CPRM2226FPM, CPRM2226FPWM, CPRM2226M; 4. Small Adult Resuscitator REF #s: CPRM4416FPMC, CPRM4416M, CPRM4426M.
Publicado el 29 de noviembre de 2023 · Inicio del retiro: 11 de octubre de 2023
Resumen
FDA registra el retiro Z-0369-2024 de dispositivos médicos de Clase II, publicado el 29 de noviembre de 2023 e iniciado el 11 de octubre de 2023. Empresa responsable: SunMed Holdings, LLC; Estado: En curso; Cantidad reportada: N/A; Lote: UDI-DI:10888277628298, (Case UDI: 20888277698885, ITEM # CPRM1116FM); UDI-DI:10888277667785, (Case UDI: 20888277667782, ITEM # CPRM1116FMT); UDI-DI:10888277654815, (Case UDI: 20888277654812, ITEM # CPRM1116FPM); UDI-DI:10193489082685, (Case UDI: 20193489021544, ITEM # CPRM1116FPM4); UDI-DI:10888277628304, (Case UDI: 20888277628301, ITEM # CPRM1116M); UDI-DI:10888277698826, (Case UDI: 20888277698823, ITEM # CPRM1116MC); UDI-DI:10888277628311, (Case UDI: 20888277628318, ITEM # CPRM1116PM); UDI-DI:10888277674004, (Case UDI: 20888277674001, ITEM # CPRM1116PMT); UDI-DI:10888277629677, (Case UDI: 20888277629674, ITEM # CPRM1126FM); UDI-DI:10884389810293, (Case UDI: 20884389810290, ITEM # CPRM1126FPM); UDI-DI:10193489001945, (Case UDI: 20193489001942, ITEM # CPRM1126FPWM); UDI-DI:10888277628328, (Case UDI: 20888277628325, ITEM # CPRM1126M); UDI-DI:10888277655140, (Case UDI: 20888277655147, ITEM # CPRM2216FM); UDI-DI:10888277698857, (Case UDI: 20888277717678, ITEM # CPRM2216FM3); UDI-DI:10888277632332, (Case UDI: 20888277632339, ITEM # CPRM2216FPM); UDI-DI:10193489001150, (Case UDI: 20193489001157, ITEM # CPRM2216FPM3); UDI-DI:10193489181937, (Case UDI: 20193489181934, ITEM # CPRM2216FPMC); UDI-DI:10888277632349, (Case UDI: 20888277632346, ITEM # CPRM2216M); UDI-DI:10193489009750, (Case UDI: 20193489009757, ITEM # CPRM2216P3M3); UDI-DI:10888277669833, (Case UDI: 20888277669830, ITEM # CPRM2216PCM); UDI-DI:10888277629684, (Case UDI: 20888277629681, ITEM # CPRM2216PM); UDI-DI:10193489016000, (Case UDI: 40884389121260, ITEM # CPRM2216PM6); UDI-DI:10888277655157, (Case UDI: 20888277655154, ITEM # CPRM2226FM); UDI-DI:10888277659766, (Case UDI: 20888277659763, ITEM # CPRM2226FM3); UDI-DI:10884389811337, (Case UDI: 20884389811334, ITEM # CPRM2226FPM); UDI-DI:10193489001952, (Case UDI: 20193489001959, ITEM # CPRM2226FPWM); UDI-DI:10888277632356, (Case UDI: 20888277632353, ITEM # CPRM2226M); UDI-DI:10888277656130, (Case UDI: 20888277656137, ITEM # CPRM3312FM); UDI-DI:10888277698871, (Case UDI: 20888277698878, ITEM # CPRM3312FM123); UDI-DI:10888277654822, (Case UDI: 20888277654829, ITEM # CPRM3312FPM); UDI-DI:10193489001167, (Case UDI: 20193489001164, ITEM # CPRM3312FPM3); UDI-DI:10193489181944, (Case UDI: 20193489181941, ITEM # CPRM3312FPMC); UDI-DI:10888277632363, (Case UDI: 20888277632360, ITEM # CPRM3312M); UDI-DI:10888277698888, (Case UDI: 40888277362605, ITEM # CPRM3312M3); UDI-DI:10193489009767, (Case UDI: 20193489009764, ITEM # CPRM3312P3M3); UDI-DI:10888277718652, (Case UDI: 20888277718569, ITEM # CPRM3312PM); UDI-DI:10884389975893, (Case UDI: 20884389975890, ITEM # CPRM3312PM2); UDI-DI:10888277675193, (Case UDI: 20888277675190, ITEM # CPRM3312PMC2); UDI-DI:10888277674028, (Case UDI: 20888277674025, ITEM # CPRM3312PMTD1); UDI-DI:10884389975909, (Case UDI: 20884389975906, ITEM # CPRM3312PMX); UDI-DI:10888277667815 (Case UDI: 20888277667812, ITEM # CPRM3312PWM); UDI-DI:10080196659777, (Case UDI: 20080196659774, ITEM # CPRM3322FM); UDI-DI:10888277632370, (Case UDI: 20888277632377, ITEM # CPRM3322M); UDI-DI:10888277718562, (Case UDI: 20888277718569, ITEM # CPRM3322PM); UDI-DI:10193489181951, (Case UDI: 20193489181958, ITEM # CPRM4416FPMC); UDI-DI:10888277700499, (Case UDI: 20888277700496, ITEM # CPRM4416M); UDI-DI:10888277700512, (Case UDI: 20888277700519, ITEM # CPRM4426M). All lots manufactured from 9/21/2018 to present. Latest expiration Date 10/11/2028. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
Qué hacer
Voluntary: Firm initiated · E-Mail
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- SunMed Holdings, LLC
- Distribuidor
- US Nationwide distribution including in the states of AR, AZ, CA®CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, NC, ND, NH, NJ, NM, NV, NY, OH, SC, TN, TX, UT, VA, & WV.
- Lote
- UDI-DI:10888277628298, (Case UDI: 20888277698885, ITEM # CPRM1116FM); UDI-DI:10888277667785, (Case UDI: 20888277667782, ITEM # CPRM1116FMT); UDI-DI:10888277654815, (Case UDI: 20888277654812, ITEM # CPRM1116FPM); UDI-DI:10193489082685, (Case UDI: 20193489021544, ITEM # CPRM1116FPM4); UDI-DI:10888277628304, (Case UDI: 20888277628301, ITEM # CPRM1116M); UDI-DI:10888277698826, (Case UDI: 20888277698823, ITEM # CPRM1116MC); UDI-DI:10888277628311, (Case UDI: 20888277628318, ITEM # CPRM1116PM); UDI-DI:10888277674004, (Case UDI: 20888277674001, ITEM # CPRM1116PMT); UDI-DI:10888277629677, (Case UDI: 20888277629674, ITEM # CPRM1126FM); UDI-DI:10884389810293, (Case UDI: 20884389810290, ITEM # CPRM1126FPM); UDI-DI:10193489001945, (Case UDI: 20193489001942, ITEM # CPRM1126FPWM); UDI-DI:10888277628328, (Case UDI: 20888277628325, ITEM # CPRM1126M); UDI-DI:10888277655140, (Case UDI: 20888277655147, ITEM # CPRM2216FM); UDI-DI:10888277698857, (Case UDI: 20888277717678, ITEM # CPRM2216FM3); UDI-DI:10888277632332, (Case UDI: 20888277632339, ITEM # CPRM2216FPM); UDI-DI:10193489001150, (Case UDI: 20193489001157, ITEM # CPRM2216FPM3); UDI-DI:10193489181937, (Case UDI: 20193489181934, ITEM # CPRM2216FPMC); UDI-DI:10888277632349, (Case UDI: 20888277632346, ITEM # CPRM2216M); UDI-DI:10193489009750, (Case UDI: 20193489009757, ITEM # CPRM2216P3M3); UDI-DI:10888277669833, (Case UDI: 20888277669830, ITEM # CPRM2216PCM); UDI-DI:10888277629684, (Case UDI: 20888277629681, ITEM # CPRM2216PM); UDI-DI:10193489016000, (Case UDI: 40884389121260, ITEM # CPRM2216PM6); UDI-DI:10888277655157, (Case UDI: 20888277655154, ITEM # CPRM2226FM); UDI-DI:10888277659766, (Case UDI: 20888277659763, ITEM # CPRM2226FM3); UDI-DI:10884389811337, (Case UDI: 20884389811334, ITEM # CPRM2226FPM); UDI-DI:10193489001952, (Case UDI: 20193489001959, ITEM # CPRM2226FPWM); UDI-DI:10888277632356, (Case UDI: 20888277632353, ITEM # CPRM2226M); UDI-DI:10888277656130, (Case UDI: 20888277656137, ITEM # CPRM3312FM); UDI-DI:10888277698871, (Case UDI: 20888277698878, ITEM # CPRM3312FM123); UDI-DI:10888277654822, (Case UDI: 20888277654829, ITEM # CPRM3312FPM); UDI-DI:10193489001167, (Case UDI: 20193489001164, ITEM # CPRM3312FPM3); UDI-DI:10193489181944, (Case UDI: 20193489181941, ITEM # CPRM3312FPMC); UDI-DI:10888277632363, (Case UDI: 20888277632360, ITEM # CPRM3312M); UDI-DI:10888277698888, (Case UDI: 40888277362605, ITEM # CPRM3312M3); UDI-DI:10193489009767, (Case UDI: 20193489009764, ITEM # CPRM3312P3M3); UDI-DI:10888277718652, (Case UDI: 20888277718569, ITEM # CPRM3312PM); UDI-DI:10884389975893, (Case UDI: 20884389975890, ITEM # CPRM3312PM2); UDI-DI:10888277675193, (Case UDI: 20888277675190, ITEM # CPRM3312PMC2); UDI-DI:10888277674028, (Case UDI: 20888277674025, ITEM # CPRM3312PMTD1); UDI-DI:10884389975909, (Case UDI: 20884389975906, ITEM # CPRM3312PMX); UDI-DI:10888277667815 (Case UDI: 20888277667812, ITEM # CPRM3312PWM); UDI-DI:10080196659777, (Case UDI: 20080196659774, ITEM # CPRM3322FM); UDI-DI:10888277632370, (Case UDI: 20888277632377, ITEM # CPRM3322M); UDI-DI:10888277718562, (Case UDI: 20888277718569, ITEM # CPRM3322PM); UDI-DI:10193489181951, (Case UDI: 20193489181958, ITEM # CPRM4416FPMC); UDI-DI:10888277700499, (Case UDI: 20888277700496, ITEM # CPRM4416M); UDI-DI:10888277700512, (Case UDI: 20888277700519, ITEM # CPRM4426M). All lots manufactured from 9/21/2018 to present. Latest expiration Date 10/11/2028.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAVentlab, LLC AirFlow manual resuscitator devices with integrated manometer, multiple accessory configurations29 de noviembre de 2023
- FDACompress Device Segmental Anchor Plug, 18 MM-For Correction of revision of unsuccessful osteotomy,29 de noviembre de 2023
- FDAGEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007,29 de noviembre de 2023
- FDACompress Device Segmental Anchor Plug, 20 MM-For Correction of revision of unsuccessful osteotomy,29 de noviembre de 2023
- FDAstryker Sage PrimaFit External Urine Management System for the Female Anatomy29 de noviembre de 2023
- FDAAir-Q3 AirQ3 Standard ILA with Manual Cuff Inflation and GA Size REF Number 0.0 50005 0.5 50055 1.0 50105 1.51 de febrero de 2023
Preguntas sobre este retiro
¿Medline manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Adult Resuscitator REF #s: CPRM1116FPM, CPRM1116FPM4, CPRM1116M, CPRM1116MC, CPRM1116PM, CPRM1116PMT, CPRM1126FM, CPRM1126FPM, CPRM1126FPWM, CPRM1126M; 2. Infant Resuscitator REF #s: CPRM3312FM, CPRM3312FM123, CPRM3312FPM, CPRM3312FPM3, CPRM3312FPMC, CPRM3312M, CPRM3312M3, CPRM3312P3M3, CPRM3312PM, CPRM3312PM2, CPRM3312PMC2, CPRM3312PMTD1, CPRM3312PMX, CPRM3312PWM, CPRM3322FM, CPRM3322M, CPRM3322PM; 3. Pediatric Resuscitator REF #s: CPRM2216FM, CPRM2216FM3, CPRM2216FPM, CPRM2216FPM3, CPRM2216FPMC, CPRM2216M, CPRM2216P3M3, CPRM2216PCM, CPRM2216PM, CPRM2216PM6, CPRM2226FM, CPRM2226FM3, CPRM2226FPM, CPRM2226FPWM, CPRM2226M; 4. Small Adult Resuscitator REF #s: CPRM4416FPMC, CPRM4416M, CPRM4426M. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0369-2024 el 29 de noviembre de 2023. Peligro reportado: A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · E-Mail