GE Healthcare - Giraffe and Panda Bag and Mask Resuscitation System;Giraffe Stand-alone Infant Resuscitation System. Product Usage: The Bag and Mask Resuscitation System provides the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant. These are clinical practices that represent the established standard of care.
Publicado el 5 de marzo de 2014 · Inicio del retiro: 1 de octubre de 2013
Resumen
FDA registra el retiro Z-1071-2014 de dispositivos médicos de Clase I, publicado el 5 de marzo de 2014 e iniciado el 1 de octubre de 2013. Empresa responsable: GE Healthcare, LLC; Estado: Finalizado; Cantidad reportada: 12, 373 units total, 8,807 = US, 3,394 = OUS, 172 = unknown; Lote: QABS50026, QABR50066, QABR50068, QABR50133, QABS50023, QABR50102, QABR50103, QABR50064, QABR50065, QABR50067, QABS50070, QABS50069, QABR50275, QABS50027, QABR50274. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- GE Healthcare, LLC
- Distribuidor
- Worldwide Distribution - USA Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL,GA. HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and the countries of Albania, Austria, Australia, Bahrain, Belarus, Belgium, Canada, China, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Greece, Guatemala, Germany, Honduras, Hong Kong, Hungry, India, Indonesia, Ireland, Italy, Kazakhstan, Korea, Lithuania, Malaysia, Mexico, Netherlands, Norway, Peru, Poland, Romania, Russia, Saudi Arabia, South Africia, Spain, Sweden, Switzerland Turkey, United Arab, United Kingdom, and Venezuela.
- Lote
- QABS50026, QABR50066, QABR50068, QABR50133, QABS50023, QABR50102, QABR50103, QABR50064, QABR50065, QABR50067, QABS50070, QABS50069, QABR50275, QABS50027, QABR50274
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACentricity PACS Software Version 7.0 SP0.0.4.73 de agosto de 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.27 de julio de 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images and13 de julio de 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 20196 de julio de 2022
- FDAGE Healthcare - Giraffe Warmer;Panda Warmer;Panda iRes Warmer, Giraffe Stand-alone Infant Resuscitation5 de marzo de 2014
- FDAGE Healthcare - Giraffe® and Panda® iRes Infant Warmer Product Usage: The infant radiant warmers provide5 de marzo de 2014
Preguntas sobre este retiro
¿GE Healthcare - Giraffe and Panda Bag and Mask Resuscitation System;Giraffe Stand-alone Infant Resuscitation System. Product Usage: The Bag and Mask Resuscitation System provides the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant. These are clinical practices that represent the established standard of care. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1071-2014 el 5 de marzo de 2014. Peligro reportado: Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter