FDAFDA · Alimentos y medicamentos·N.º oficial Z-1081-2014
RT Desktop The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as the user interface for the SL/SLi Series Linear Accelerators that are used for the radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.
Publicado el 12 de marzo de 2014 · Inicio del retiro: 25 de noviembre de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Elekta, Inc.
- Distribuidor
- Worldwide Distribution - USA including AL, CA, CT, GA, FL, IL, IN, KS, KY, LA, MD, MA, MI, MN, MO, NJ, NY, OH, PA, TN, TX, VA, and Internationally to Argentina, Austria, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cuba, Czech Republic, Egypt, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Lebanon, Mexico, Morocco, Netherland, Nigeria, Pakistan, Panama, Peru, Philippines, Portugal, Russia, Serbia, Solvenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, United Kingdom, Venezuela, Vietnam, and Yemen.
- Lote
- R4.2, R5.0, R5.1, R6.0, R6.1
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States