Elekta Unity, Image-Guided Radiation Therapy System Product Usage: Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with compatible Treatment Planning and Oncology Information Systems.
Publicado el 17 de abril de 2019 · Inicio del retiro: 12 de marzo de 2019
Resumen
FDA registra el retiro Z-1109-2019 de dispositivos médicos de Clase II, publicado el 17 de abril de 2019 e iniciado el 12 de marzo de 2019. Empresa responsable: Elekta Limited; Estado: Finalizado; Cantidad reportada: 12 units; Lote: Unity Systems: 13553-006/600003, 13371-002/600021, 10817-011/600010, 10562-011/600016, 10719-T03001/600013, 30004070-001/600023, 10420-AVL-U/600008, 11014-UMCU-U/600007, 10157-006/600014, 11611-40/600011, 30001461-J001/600002, 11489-21/600009,. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Users need to be aware when using these protocols for daily on line plan adaptation that: 1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion cycle. The images are based on data acquired around full expiration. 2) The display of the images in the Elekta Unity Application software does not provide information about the protocol used to acquire the image eg. with or without respiratory triggering. Users must select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan.
Qué hacer
Voluntary: Firm initiated · E-Mail
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Elekta Limited
- Distribuidor
- Worldwide Distribution - US Nationwide in the states of TX, NY, WI, and Denmark, Germany, Italy, Netherlands, Sweden, United Kingdom
- Lote
- Unity Systems: 13553-006/600003, 13371-002/600021, 10817-011/600010, 10562-011/600016, 10719-T03001/600013, 30004070-001/600023, 10420-AVL-U/600008, 11014-UMCU-U/600007, 10157-006/600014, 11611-40/600011, 30001461-J001/600002, 11489-21/600009,
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United Kingdom
Retiros relacionados
- FDADivalproex Sodium Extended-Release Tablets, USP, 250 mg a) 100-count bottle (NDC 55111-0533-01), b) 500-count24 de abril de 2019
- FDAVolumex (Iodinated I 131 Albumin) Injection USP, 25 uCi per 1 mL syringe, Rx Only, Manufactured for Daxor24 de abril de 2019
- FDAiTotal CR Total Knee Replacement System, TCR-121-1111 (Catalog No.) / TCR-121-1111-010102 (Model No.)17 de abril de 2019
- FDAGravity IV Set with 3-port Closed Stopcock Manifold Model CM42500E-0717 de abril de 2019
- FDAWalgreens Probiotic 12.5 Billion Active Cultures, 30 capsules; Dietary Supplement. Probiotic Blend 12.517 de abril de 2019
- FDAImpossible Foods Bulk, 5 lb. vacuum sealed plastic-pack. Keep Frozen; This is a plant product that is looks17 de abril de 2019
Preguntas sobre este retiro
¿Elekta Unity, Image-Guided Radiation Therapy System Product Usage: Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with compatible Treatment Planning and Oncology Information Systems. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1109-2019 el 17 de abril de 2019. Peligro reportado: Users need to be aware when using these protocols for daily on line plan adaptation that: 1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion cycle. The images are based on data acquired around full expiration. 2) The display of the images in the Elekta Unity Application software does not provide information about the protocol used to acquire the image eg. with or without respiratory triggering. Users must select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · E-Mail