BIOFIRE Blood Culture Identification 2 (BCID2) Panel when used with: BD BACTEC Lytic Anaerobic medium BD BACTEC Peds Plus medium BD BACTEC Plus Aerobic medium BD BACTEC Plus Anaerobic medium BD BACTEC Standard Aerobic medium BD BACTEC Standard Anaerobic medium The BIOFIRE Blood Culture Identification 2 (BCID2) Panel is a multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures.
Publicado el 21 de febrero de 2024 · Inicio del retiro: 26 de enero de 2024
Resumen
FDA registra el retiro Z-1139-2024 de dispositivos médicos de Clase II, publicado el 21 de febrero de 2024 e iniciado el 26 de enero de 2024. Empresa responsable: BioFire Diagnostics, LLC; Estado: En curso; Cantidad reportada: 32,910 panels/pouches; Lote: For BioFire Blood Culture Identification 2 (BDID2) Panel Part Number: RFIT-ASY-0147 UDI-DI code: 00815381020338 Lot Numbers: All non-expired BCID2 Panel lots, if used with the BD BACTEC vials including but not limited to, the bottle types listed in Table 1. NOTE: The affected blood culture vial types most likely contain non-viable Candida tropicalis nucleic acid/DNA fragments that can be detected by the BCID2 Panel. Therefore, all lots of the BioFire BCID2 Panel would be affected if they are used with the affected blood culture vials. Neither the BioFire BCID2 test kits nor the BD BACTEC vials have a defect - both products are performing as intended. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Due to a trend in false positive Candida tropicalis results on blood culture panel when using certain BD BACTEC vial mediums
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- BioFire Diagnostics, LLC
- Distribuidor
- Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Andorra, United Arab Emirates, Armenia, Argentina, Austria, Australia, Bosnia, Belgium, Bulgaria, Bahrain, Brunei, Bolivia, Brazil, Canada, Switzerland, Cote D'Ivoire, Chile, China, Colombia, Costa Rica, Cyprus, Cech Republic, Germany, Denmark, Dominican Republic, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, France, Gabon, United Kingdom, Georgia, Guadeloupe, Greece, Guatemala, Hong Kong, Honduras, Croatia, Hungary, Iceland, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, Korea, Kuwait, Lebanon, Lithuania, Latvia, Morocco, Macedonia, Mali, Mongolia, Macau, Mexico, Malaysia, New Caledonia, Nigeria, Nicaragua, Netherlands, Norway, New Zealand, Oman, Panama, Peru, French Polynesia, Philippines, Pakistan, Poland, St. Pierre/Miquelon, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, El Salvador, Thailand, Tunisia, Turkey, Taiwan, Uganda, Uruguay, Vatican City, Venezuela, Vietnam, South Africa, and Zimbabwe.
- Lote
- For BioFire Blood Culture Identification 2 (BDID2) Panel Part Number: RFIT-ASY-0147 UDI-DI code: 00815381020338 Lot Numbers: All non-expired BCID2 Panel lots, if used with the BD BACTEC vials including but not limited to, the bottle types listed in Table 1. NOTE: The affected blood culture vial types most likely contain non-viable Candida tropicalis nucleic acid/DNA fragments that can be detected by the BCID2 Panel. Therefore, all lots of the BioFire BCID2 Panel would be affected if they are used with the affected blood culture vials. Neither the BioFire BCID2 test kits nor the BD BACTEC vials have a defect - both products are performing as intended.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAGeneric Name: FilmArray® Pneumonia Panel Generic Name: FilmArray® Pneumonia Panel Model/Catalog Number:21 de octubre de 2026
- FDAGeneric Name: FilmArray® Pneumonia Panel plus Generic Name: FilmArray® Pneumonia Panel plus Model/Catalog21 de octubre de 2026
- FDAGeneric Name: BioFire® Joint Infection (JI) Panel Generic Name: BioFire® Joint Infection (JI) Panel21 de octubre de 2026
- FDABIOFIRE Respiratory Panel 2.1 (RP2.1) REF 423742 UDI code: 00815381020529 PCR-based multiplexed nucleic acid19 de agosto de 2026
- FDAglobal Medium, Model Numbers LGGG-020 (20mL bottle), LGGG-050 (50mL bottle), LGGG-100 (100mL bottle). Intended21 de febrero de 2024
- FDAFilmArray Gastrointestinal (GI) Panel RFIT-ASY-0116 (30 test kit) RFIT-ASY-0104 (6 test kit) Gastrointestinal21 de febrero de 2024
Preguntas sobre este retiro
¿BIOFIRE Blood Culture Identification 2 (BCID2) Panel when used with: BD BACTEC Lytic Anaerobic medium BD BACTEC Peds Plus medium BD BACTEC Plus Aerobic medium BD BACTEC Plus Anaerobic medium BD BACTEC Standard Aerobic medium BD BACTEC Standard Anaerobic medium The BIOFIRE Blood Culture Identification 2 (BCID2) Panel is a multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1139-2024 el 21 de febrero de 2024. Peligro reportado: Due to a trend in false positive Candida tropicalis results on blood culture panel when using certain BD BACTEC vial mediums
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter