Panaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can also set independent alarm limits for data send by the bedside monitor.

Publicado el 19 de marzo de 2014 · Inicio del retiro: 8 de enero de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Mindray has identified two issues with the Panorama Central Station. Full Disclosure data may lag behind the Panorama system time clock when the Panorama Central Station is in use with Mindray¿s V-Series Monitor. Additionally, event indicators lag behind the Panorama system time clock when the Panorama Central Station is in use with a Panorama Telepack.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Mindray DS USA, Inc. dba Mindray North America
Distribuidor
Worldwide Distribution - USA Nationwide Distribution
Lote
p/n 0998-00-0708-01; 0998-UC-0708-01 and 0020-00-0205-XXX
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States