Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1177-2013

Maquet Getinge Group battery modules for use with Maquet SERVO-i ventilator systems (Part #6487180). Battery for use with the SERVO-i Ventilator System. The SERVO-i Ventilator System is intended for treating and monitoring patients ranging from neonates to adults with respiratory failure or insufficiency. The ventilator is equipped with at least 2, NiMH battery modules (Part No. 6487180) which automatically supply 12V DC Power in case of an AC power failure, ensuring that ventilator settings and stored data remain intact in the event of an AC power failure. The battery modules give the user a possibility to use the SERVO-i during intra-hospital transportation.

Publicado el 8 de mayo de 2013 · Inicio del retiro: 7 de marzo de 2013

Resumen

FDA registra el retiro Z-1177-2013 de dispositivos médicos de Clase I, publicado el 8 de mayo de 2013 e iniciado el 7 de marzo de 2013. Empresa responsable: Maquet Cardiovascular Us Sales, Llc; Estado: Finalizado; Cantidad reportada: 90,000 units (batteries); Lote: SERVO-i Battery Module part number is 6487180. Notice pertains to Battery Modules manufactured between February 2010 and October 2012 with date codes between 1005 and 1243. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

When using the SERVO-i on battery power, in a very small number of cases, some battery modules distributed after January 31, 2010 have a shorter battery run time than expected.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Maquet Cardiovascular Us Sales, Llc
Distribuidor
Worldwide distribution: USA (nationwide) and countries of: ALGERIA, ANDORRA, ANTILLES (Netherlands), ARGENTINA, ARUBA, AUSTRALIA, AUSTRIA, AZERBADJAN, BANGLADESH, BELARUS, BELGIUM, BOLIVIA, BOTSWANA, BRAZIL, BRUNEI, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CONGO, COSTA RICA, CROATIA, CUBA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, YUGOSLAVIA (Former), FIJI, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, GUATEMALA, GUYANA, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, KENYA, KUWAIT, KYRGYZSTAN, LATVIA, LIBYA, LITHUANIA, LUXEMBURG, MACAU, MACEDONIA, MALAYSIA, MALDIVES, MARTINIQUE (French), MEXICO, MOLDAVIA, MOROCCO, NEPAL, NETHERLANDS, NEW CALEDONIA (French), NEW ZEALAND, NICARAGUA, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PANAMA, PAPUA NEW GUINEA, PARAGUAY, PERU, PHILIPPINES, POLAND, POLYNESIA (French), PORTUGAL, QATAR, REUNION (French), ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SEYCHELLES, SINGAPORE, SLOVAK REPUBLIC, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SUDAN, SURINAM, SWEDEN, SWITZERLAND, SYRIA, TADJIKISTAN, TAIWAN, TANZANIA, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VATICAN city state, VENEZUELA, VIETNAM and YEMEN.
Lote
SERVO-i Battery Module part number is 6487180. Notice pertains to Battery Modules manufactured between February 2010 and October 2012 with date codes between 1005 and 1243.
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Maquet Getinge Group battery modules for use with Maquet SERVO-i ventilator systems (Part #6487180). Battery for use with the SERVO-i Ventilator System. The SERVO-i Ventilator System is intended for treating and monitoring patients ranging from neonates to adults with respiratory failure or insufficiency. The ventilator is equipped with at least 2, NiMH battery modules (Part No. 6487180) which automatically supply 12V DC Power in case of an AC power failure, ensuring that ventilator settings and stored data remain intact in the event of an AC power failure. The battery modules give the user a possibility to use the SERVO-i during intra-hospital transportation. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1177-2013 el 8 de mayo de 2013. Peligro reportado: When using the SERVO-i on battery power, in a very small number of cases, some battery modules distributed after January 31, 2010 have a shorter battery run time than expected.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter