Transport Stretcher, and Head and Neck Surgery Stretcher Non-powered stretcher
Publicado el 23 de marzo de 2016 · Inicio del retiro: 5 de febrero de 2016
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The hydraulic jack assemblies (PN: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) were manufactured using the incorrect oil. The oil used does not contain the same wear resistance additives as the specified oils and therefore its use increases the potential for wear which may cause a leak event or jack drift. Testing was done to evaluate the jacks with incorrect oil and leak events we
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Stryker Medical Division of Stryker Corporation
- Distribuidor
- Distributed in DC and the states of AL, CA, FL, MN, MT, NC, NE, NY, PA, RI, TN, TX, UT, and WV.
- Lote
- The hydraulic jack assemblies (Part numbers: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) used as service parts on Stretchers assembled from 8/28/15 - 8/29/15
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States