Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1197-2013

Presource PBDS, Knee Arthroscopy, Kit, Circulator

Publicado el 15 de mayo de 2013 · Inicio del retiro: 26 de marzo de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Various Presource Kits contain a pre-assembled Filter and Anesthesia Circuit that may contain outer plastic packaging on one or more components. If the packaging is removed without disassembling the components, remnants of the plastic from the packaging material may become lodged in the filter potentially causing an obstruction of airflow.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Cardinal Health, Medical Products & Services
Distribuidor
Distributed in the states of AR, MS, SC, and TN.
Lote
Catalog Number: PO24AKNPC01; Lots: 851288, 877594, 901672, 917786, 948443, 972424
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States