FDAFDA · Alimentos y medicamentos·N.º oficial Z-1200-2013
Presource PBDS, Shoulder Arthroscopy, Kit, Circulator
Publicado el 15 de mayo de 2013 · Inicio del retiro: 26 de marzo de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Various Presource Kits contain a pre-assembled Filter and Anesthesia Circuit that may contain outer plastic packaging on one or more components. If the packaging is removed without disassembling the components, remnants of the plastic from the packaging material may become lodged in the filter potentially causing an obstruction of airflow.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Cardinal Health, Medical Products & Services
- Distribuidor
- Distributed in the states of AR, MS, SC, and TN.
- Lote
- Catalog Number: PO24SAGWF01; Lots: 847129, 877843, 905183, 931801, 935180, 946651, 957422, 969585
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.