Elscint APEX Helix, dual-head, multi-purpose, Slip-Ring, digital gamma camera Product Usage: The intended use of the Quasar system is to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition typed and optional imaging features designed to enhance image quality in Oncology, Cardiology, Neurology and other clinical diagnostic imaging applications. The scanning modes include planar (Static, Multi-gated, Dynamic, Whole body scanning) and tomographic (SPECT, Gated SPECT, Whole body SPECT, Camera based PET). Acquisition types include single and multi-isotope/multi-peak frame/list mode single-photon and positron imaging. Optional imaging-enhancement features include assortment of collimators, gating by physiological signals, real-time automatic body contouring, and CT-based attenuation correction and functional anatomic mapping.
Publicado el 19 de marzo de 2014 · Inicio del retiro: 7 de febrero de 2014
Resumen
FDA registra el retiro Z-1200-2014 de dispositivos médicos de Clase II, publicado el 19 de marzo de 2014 e iniciado el 7 de febrero de 2014. Empresa responsable: GE Healthcare, LLC; Estado: Finalizado; Cantidad reportada: 3062 units (1182 US, 1880 OUS); Lote: all serial numbers are listed in Product 1. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Potential safety issue related to unintended radial detector motion, which may occur during patient setup or at the end of a patient scan. This failure mode has been observed on a single system, in which the detector moved to the home position and then slowly starting drifting inward toward the patient. This generated both a visual and audio alert condition on both the console and the gantry. N
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- GE Healthcare, LLC
- Distribuidor
- Worldwide Distribution-USA (nationwide) including Puerto Rico and Guam and the countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Korea (Republic of), Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Malta, Martinique, Mexico, Morocco, Namibia, New Zealand, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, and United Kingdom.
- Lote
- all serial numbers are listed in Product 1.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACentricity PACS Software Version 7.0 SP0.0.4.73 de agosto de 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.27 de julio de 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images and13 de julio de 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 20196 de julio de 2022
- FDAGE Healthcare Varicam/VG and VG Hawkeye devices Product Usage: The intended use of the Quasar system is to19 de marzo de 2014
- FDAMar Cor Purification CWP Reverse Osmosis System. Models 101, 102, 104, and 106.19 de marzo de 2014
Preguntas sobre este retiro
¿Elscint APEX Helix, dual-head, multi-purpose, Slip-Ring, digital gamma camera Product Usage: The intended use of the Quasar system is to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition typed and optional imaging features designed to enhance image quality in Oncology, Cardiology, Neurology and other clinical diagnostic imaging applications. The scanning modes include planar (Static, Multi-gated, Dynamic, Whole body scanning) and tomographic (SPECT, Gated SPECT, Whole body SPECT, Camera based PET). Acquisition types include single and multi-isotope/multi-peak frame/list mode single-photon and positron imaging. Optional imaging-enhancement features include assortment of collimators, gating by physiological signals, real-time automatic body contouring, and CT-based attenuation correction and functional anatomic mapping. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1200-2014 el 19 de marzo de 2014. Peligro reportado: Potential safety issue related to unintended radial detector motion, which may occur during patient setup or at the end of a patient scan. This failure mode has been observed on a single system, in which the detector moved to the home position and then slowly starting drifting inward toward the patient. This generated both a visual and audio alert condition on both the console and the gantry. N
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter