FDAFDA · Alimentos y medicamentos·N.º oficial Z-1217-2014
Gibralt Spine System Polyaxial Screw Intended to promote fusion of the cervical spine, and the thoracic spine, (C3-T3), the Gibralt Spinal System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C3-T3) spine.
Publicado el 26 de marzo de 2014 · Inicio del retiro: 6 de enero de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The products have been identified to contain springs which have been manufactured with type-304 stainless steel rather than the specified titanium alloy.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Exactech, Inc.
- Distribuidor
- USA Distribution including the states of FL, NY, OR, IL, PA, MO and NJ.
- Lote
- size 3.5mm x 10mm, Catalog #05-000-20-3510, Lot Number #35917001. size 3.5mm x 14mm, Catalog #05-000-20-3514, Lot Number 35917002 and 40023002. size 3.5mm x 20mm, Catalog #05-000-20-3520, Lot Number 35917003. size 3.5mm x 26mm, Catalog #05-000-20-3526, Lot Number 35917004. size 4.0mm x 14mm, Catalog #05-000-20-4014, Lot Number 35917005. size 4.0mm x 20mm, Catalog #05-000-20-4020, Lot Number 35917006. size 4.0mm x 26mm, Catalog #05-000-20-4026, Lot Number 35917007. 1/4 Shank, size 3.5mm x 26mm, Catalog #05-000-20-3526, Lot Number 35917008. 1/4 Shank, size 3.5mm x 30mm, Catalog #05-000-20-3530, Lot Number 35917009. 1/4 Shank, size 4.0mm x 26mm, Catalog #05-000-20-4026, Lot Number 35917010. 1/4 Shank, size 4.0mm x 30mm, Catalog #05-000-20-4030, Lot Number 35917011.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States