Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1226-2014

VITROS Chemistry Products White Reference Slides, White Correction Factor Slide (WCF), DT Slides, 25 Slides, Part Number J02315; Made in USA by Ortho- Clinical Diagnostics, Inc. 100 Indigo Creek Drive, Rochester, NY 14626 --- This product is a part/ tool used when performing the correction factors adjustment on the VITROS Chemistry Systems. It does not have a 510(k) number. Because the WCF slide is a part/ tool, and is not a device used for diagnostic testing, there is no Intended Use statement. The following summary is provided as a description/ explanation of the WCF slide. The WCF slide is a part/ tool used in the reflectometer correction factor adjustment procedure in order to optimize the optical calibration of the VITROS 250/350/5,1 FS/4600 Chemistry Systems and the VITROS 5600 Integrated

Publicado el 26 de marzo de 2014 · Inicio del retiro: 9 de enero de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

An incorrect assay value (0.3785) was printed on the product carton for the 460 nm wavelength (filter) that may cause a delay in the ability to obtain and report TBIL, Bu, Bc, and derived tests results. The correct assay value is 0.8658.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Ortho-Clinical Diagnostics
Distribuidor
Worldwide Distribution - USA (nationwide) Puerto Rico, and Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain, and Venezuela.
Lote
Lot 9052-0045-6358, expiry date 01 Aug 2015
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta