Discovery NM/CT 670 DR Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce:

Publicado el 8 de mayo de 2019 · Inicio del retiro: 20 de marzo de 2019

Resumen

FDA registra el retiro Z-1259-2019 de dispositivos médicos de Clase II, publicado el 8 de mayo de 2019 e iniciado el 20 de marzo de 2019. Empresa responsable: GE Healthcare, LLC; Estado: Finalizado; Cantidad reportada: 11 units; Lote: Mfg. Lot or Serial # System ID Model, Catalogue or Code # DRGX56002 SHIPPED (GON 4683652) H3100BT DDRX34058 909558D670NM H3100AN DDRX34056 650723D670 H3100AN DDRX34053 SHIPPED (AO 1614951) H3100AN DRGX56005 614CVA6702 H3100BT DRGX56017 110002NU06 H3100BT DRGX56004 403343D670A H3100BT DRGX56018 X7410502 H3100BT DRGX56009 083026206531618 H3100BT DRGX56012 STD01300 H3100BT DDRX34055 00362NUC04 H3100AN. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
GE Healthcare, LLC
Distribuidor
US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK
Lote
Mfg. Lot or Serial # System ID Model, Catalogue or Code # DRGX56002 SHIPPED (GON 4683652) H3100BT DDRX34058 909558D670NM H3100AN DDRX34056 650723D670 H3100AN DDRX34053 SHIPPED (AO 1614951) H3100AN DRGX56005 614CVA6702 H3100BT DRGX56017 110002NU06 H3100BT DRGX56004 403343D670A H3100BT DRGX56018 X7410502 H3100BT DRGX56009 083026206531618 H3100BT DRGX56012 STD01300 H3100BT DDRX34055 00362NUC04 H3100AN
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Discovery NM/CT 670 DR Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce: fue retirado del mercado?

Sí. FDA publicó el retiro Z-1259-2019 el 8 de mayo de 2019. Peligro reportado: During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter