NM/CT 870 CZT, Model/Catalogue/Code # H3906CW Product Usage: The GE Discovery NM/CT 670 CZT system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using a variety of scanning modes supported by various acquisition types, and a Computed Tomography system which is intended for enabling attenuation correction and anatomical localization of SPECT images and for standalone head, whole body, cardiac and vascular X-ray Computed Tomography applications.

Publicado el 8 de mayo de 2019 · Inicio del retiro: 20 de marzo de 2019

Resumen

FDA registra el retiro Z-1264-2019 de dispositivos médicos de Clase II, publicado el 8 de mayo de 2019 e iniciado el 20 de marzo de 2019. Empresa responsable: GE Healthcare, LLC; Estado: Finalizado; Cantidad reportada: 16 units; Lote: Mfg. Lot or Serial # System ID 87CY63017 SHIPPED (AO 1630621) 87CX63007 AH0101NM05 87CX63014 030078NU28 87CX63008 190020NU01 87CX63006 030240NU11 87CX63015 705743NM870 87CX63011 M4184879 87CX63003 M9886873 87CX63004 X171582003 87CX63009 M171582018 87CX63010 UC1796NU01 87CX63013 NO1068MN18 87CX63005 A004NU20 87CX63002 C028NU01 87CX63012 C028NU02 87CX63001 900010NU12. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
GE Healthcare, LLC
Distribuidor
US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK
Lote
Mfg. Lot or Serial # System ID 87CY63017 SHIPPED (AO 1630621) 87CX63007 AH0101NM05 87CX63014 030078NU28 87CX63008 190020NU01 87CX63006 030240NU11 87CX63015 705743NM870 87CX63011 M4184879 87CX63003 M9886873 87CX63004 X171582003 87CX63009 M171582018 87CX63010 UC1796NU01 87CX63013 NO1068MN18 87CX63005 A004NU20 87CX63002 C028NU01 87CX63012 C028NU02 87CX63001 900010NU12
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿NM/CT 870 CZT, Model/Catalogue/Code # H3906CW Product Usage: The GE Discovery NM/CT 670 CZT system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using a variety of scanning modes supported by various acquisition types, and a Computed Tomography system which is intended for enabling attenuation correction and anatomical localization of SPECT images and for standalone head, whole body, cardiac and vascular X-ray Computed Tomography applications. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1264-2019 el 8 de mayo de 2019. Peligro reportado: During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter