FDAFDA · Alimentos y medicamentos·N.º oficial Z-1302-2024
Boston Scientific POLARSHEATH Steerable Sheath 12F, REF M004CRBS3050
Publicado el 3 de abril de 2024 · Inicio del retiro: 12 de febrero de 2024
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Boston Scientific has identified a tooling error in manufacturing which may have caused delamination of the inner lumen of the sheath shaft in a subset of POLARSHEATH devices. This could result in the embolization of a fragment during use (such as flushing of the sheath or introduction of the dilator or ablation catheter).
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Boston Scientific Corporation
- Distribuidor
- Austria, Belarus, Estonia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Netherlands, Poland, Portugal, Spain, and Japan
- Lote
- GTIN 08714729992684, Batch Numbers: 32895945, 32895947, 32903274, 32903275, 32903276, 32903277, 32903278, 32903340, 32999267, 33000331, 33000333, 33000334, 33000337, 33000338, 33000339, 33052646, 33052647, 33052648, 33052649, 33061576, 33171976
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States