FDAFDA · Alimentos y medicamentos·N.º oficial Z-1302-2024

Boston Scientific POLARSHEATH Steerable Sheath 12F, REF M004CRBS3050

Publicado el 3 de abril de 2024 · Inicio del retiro: 12 de febrero de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Boston Scientific has identified a tooling error in manufacturing which may have caused delamination of the inner lumen of the sheath shaft in a subset of POLARSHEATH devices. This could result in the embolization of a fragment during use (such as flushing of the sheath or introduction of the dilator or ablation catheter).

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Boston Scientific Corporation
Distribuidor
Austria, Belarus, Estonia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Netherlands, Poland, Portugal, Spain, and Japan
Lote
GTIN 08714729992684, Batch Numbers: 32895945, 32895947, 32903274, 32903275, 32903276, 32903277, 32903278, 32903340, 32999267, 33000331, 33000333, 33000334, 33000337, 33000338, 33000339, 33052646, 33052647, 33052648, 33052649, 33061576, 33171976
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States